Reference Oscillator = Off
Active entry channel = A
Measurement units = Watts
REL mode = off
Measurement parameters (set for Sensor A and Sensor B):
Cal Factor = 100.0%
Cal Adj = 100.0%
Offset = 0.00 dB
RA Auto Range
Syntax:
RA
Remarks:
Sets the ML2430A Series to automatically select the correct range for the
measurement.
RC Recall setup
Syntax:
RC<val>EN
val:
1 to 10
Remarks:
The ML2430A Series can store up to 10 instrument configurations for con-
venient recall. The configuration parameters stored are the same parame-
ters the ML2430A Series stores in its own *SAV and *RCL commands.
Therefore, RC is equivalent to *RCL, and ST is equivalent to *SAV.
Selecting Register 0 always restores the previous power meter configura
-
tion, providing an expedient way to recover from an entry error.
RH Range hold
Syntax:
RH
Remarks:
Hold the power meter in the current range. The differences in sensor
ranges must be taken into account when the ML2430 is being used to em
-
ulate the HP 438.
RL Relative mode
Syntax:
RL<mode>
6-124 ML2430A OM
HP 438A EMULATION COMMANDS GPIB OPERATION