same command as the one which changes to AUTO averaging, it will also
update the USER averaging number.
Example:
AVG A, AUTO, 64
This command will set the system to AUTO averaging and the USER av
-
eraging number to 64. But, the Auto Averaging measurement system does
not use the USER averaging number.
NOTE
The AVG mnemonic can be sent to just change the
<mode> of averaging (MOV, REPEAT etc.) without sending
a number, but there must
be a following COMMA to indi
-
cate the <val> parameter is not being sent. See the first ex
-
ample below.
The AVG mnemonic can also be sent to just change the
User Average Number <val> (1 to 512) without changing
the averaging mode, but there must
be a COMMA to indi-
cate the <mode> parameter is not being sent. See the last
example below.
Examples:
AVG A, AUTO,
Change Sensor A to Auto Averaging (note following comma to indicate the
<val> parameter is not being sent).
AVG A, AUTO, 64
Change Sensor A to AUTO and the User Average Number to 64.
AVG A, MOV, 32
Change sensor A to MOVing average and the User Average number to
32.
AVG A, RPT,
Change sensor A to RePeaT average and keep the User Average number
as 32.
AVG A, , 128
Change sensor A User Average number to 128, but keep the previously
set averaging mode (note comma to indicate the <mode> parameter is
not being sent).
Query:
AVG? <S>
Returned
String:
AVG <S>,<MODE>,<VAL>
6-18 ML2430A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION