Remarks:
This function is used to toggle a sensor’s range hold off or on, to set a spe
-
cific range to be held, or to select AUTO ranging. RGH sent with only the
sensor parameter will toggle the sensor between holding the present oper
-
ating range and AUTO. If RGH is sent with sensor and value parameters,
the sensor range mode will be set to the range value sent.
Query:
RGH? <s>
Returned
String:
RGH <s>,<val>
RSBAUD RS232 Baud Rate
SYSTEM
Syntax:
RSBAUD <val>
val:
12,24,48,96,192 or 384 hundred bits per second
Remarks:
Sets the RS232 Baud rate for the rear panel serial port.
Query:
RSBAUD?
Returned
String:
RSBAUD <val>
RSMODE RS232 Operating Mode
SYSTEM
Syntax:
RSMODE <s>
s:
EXTCOM
SRCSWP
Remarks:
EXTCOM = External communication. GPIB commands are sent and re
-
ceived via an RS232 connection.
SRCSWP = Source sweep. Connected to a sweeper so updates to the
sweepers power of frequencies etc. are automatically updated on the
ML2430A Series also.
Query:
RSMODE?
Returned
String:
RSMODE <s>
SECURE Secure system state
SYSTEM
Syntax:
SECURE <state>
state:
ON or OFF
6-68 ML2430A OM