4-7 HOT S22 The “B_HotS22.txt” file provides prompts for setting up a Hot S
22
specifi
-
cation file. Hot S
22
is a return loss measurement of an amplifiers output
port (Port 2) while stimulus is applied to its input port. The prompts
cause the Scorpion to be placed into a state where it is ready to make cali
-
brated measurements on the users AUT. At the completion of the pro
-
gram, the state information is saved to a *.cal file. To run this program,
proceed as follows.
Step 1. Select “Tools” then “Run Calibration Procedure.”
Step 2. Double-click on “C_Hot S22.txt” to begin the Hot S
22
cal file
setup. Follow the prompt and connect open, short, and load to
Port 2 as directed.
HOT S22 OPERATIONS, CALIBRATION
4-8 ME 7840A OMM