CAL SPECIFICATION FILES TECHNICAL DETAILS
ME7840A OMM A-3
k. In each calibration specification/section, use the following structure:
3. Header:
4. “Type” Keywords, Legal Values:
a. pow Power Level Calibration
- linear cal
- flat power cal
- receiver cal
b. sp S Parameter Calibration
- 12 term
- Hot S
22
- Autocal
c. imd Instrument Setup
Sets up instrument for IMD measurements. Uses Flat Power Cal
and Receiver cal as they exist in the Scorpion's non-volatile memory.
d. harm Instrument Setup
Sets up instrument for harmonic measurement. Uses Flat Power Cal and
Receiver cal as they exist in the Scorpion's non-volatile memory.
e. powsweep Instrument Setup
Sets up instrument for IMD-mode power sweep.
Table A-1. Technical Details of Calibration Specification Files (2 of 9).
Header for section
<END>
<BEGIN>
Details of 1st cal.
CAL, 1
Details of 2nd cal.
CAL, 2
Details of Nth cal.
CAL, <N>
·
·
·
·
·
·
//comment
type, <value 1>, //comment
type, <value 1>, //comment
one or more
calibration
sequence
types
Przeglądanie stron 112