General Information 1-17 Secure Environment Workplace
Site Master User Guide PN: 10580-00252 Rev. E 1-11
Recommended Usage in a Secure Environment
Set the Site Master to save files to an external USB Flash drive:
1. Attach the external Flash drive and turn the instrument on.
2. Press the Shift button then the File (7) button.
3. Press the Save submenu key.
4. Press the Change Save Location submenu key, then select the USB drive with the
rotary knob, Up/Down arrow keys, or the touchscreen.
5. Press the Set Location submenu key.
The external USB drive is now the default location for saving files.
Note
For USB storage, Anritsu recommends P/N 2000-1520-R USB Flash Drive. Not all
after-market USB drives are compatible with the instrument. Many drives come
with a second partition that contains proprietary firmware. This partition must be
removed. Only one partition is allowed. Refer to the individual manufacturer for
instructions on how to remove it. Some drives can be made to work by reformatting
them using the FAT32 format.