5-7 Site Master Firmware Emergency Repair System Operations
5-8 PN: 10580-00252 Rev. A Site Master User Guide
8. The Firmware Update dialog will query you to confirm the process by pressing Enter to
continue or Esc to abort.
9. Select Enter and the firmware update process begins and the Firmware Update dialog
displays the following message
Updating firmware. Please Wait.
10. When complete, the instrument will reboot.
5-7 Site Master Firmware Emergency Repair
If you have problems with booting up or updating firmware on the instrument, use the
following steps to resolve the problem.
1. Press and hold the Shift, 4 (Measure), 0, and On/Off keys simultaneously until a green
bar displays at the top of the screen. The Bootstrap window opens.
2. Place the USB memory device with update firmware into the USB connector on the
instrument. Please review the MST User Guide section on copying firmware to a USB
memory device.
3. Touch the Load All command line on the screen.
4. A successful emergency repair returns:
Loading applications passed.
Loading DSP FPGA passed.
Loading SPA FPGA passed.
Loading OS passed.
5. Power down and power up the instrument.