Do not exceed the operating input power level,
voltage level, current level, and signal type that is
appropriate for the instrument being used. Refer to
your instrument’s operation manual for safe
operating practices and device limitations.
Electrostatic Discharge (ESD) can damage the highly
sensitive circuits in the instrument. ESD is most
likely to occur as test devices are being connected to,
or disconnected from, the instrument’s front and rear
panel ports and connectors. You can protect the
instrument and test devices by wearing a
static-discharge wristband. Alternatively, you can
ground yourself to discharge any static charge by
touching the outer chassis of the grounded
instrument before touching the instrument’s front
and rear panel ports and connectors. Avoid touching
the test port center conductors unless you are
properly grounded and have eliminated the
possibility of static discharge.
Repair of damage that is found to be caused by
electrostatic discharge is not covered under
warranty.
Caution
Przeglądanie stron 0
1 2 3 4 5 6 ... 469 470

Podsumowanie treści

Strona 1

Do not exceed the operating input power level,voltage level, current level, and signal type that isappropriate for the instrument being used. Refer to

Strona 2

7. Handle PCBs only by their edges.Do not handle by the edge connectors.8. Lift & handle solid state devices bytheir bodies—never by their leads.9

Strona 3 - TIP TO GET STARTED QUICKLY

4-7 MARKER KEY MENUS This key provides entry to marker functions.Menu Name Function Description GPIB MnemonicALL DISPLAYEDCHANNELSACTIVE MARKER XX Sel

Strona 4

Menu Name Function Description GPIB MnemonicALL DISPLAYEDCHANNELS(Power Sweep)ACTIVE MARKER XX Selects the active marker on all channelsMARKER TO PEAK

Strona 5

Menu Name Function Description GPIB MnemonicCHx-SXXREFERENCEPLANE X.XXXXmm(Power Sweep)SELECT DREF XX Selects the reference marker (1-12)DREF FREQXXX.

Strona 6

Menu Name Function Description GPIB MnemonicCHx-SXXREFERENCEPLANE X.XXXXmm(Power Sweep)SELECTDREF XXSelects the reference marker (1-12)DREF FREQXXX.XX

Strona 7 - Safety Symbols

Menu Name Function Description GPIB MnemonicCHx-SXXREFERENCEPLANE X.XXXXmm(Power Sweep)SELECT∆REF XXSelects the reference marker (1-12)∆REF FREQXXX.XX

Strona 8 - For Safety

Menu Name Function Description GPIB MnemonicCHx-SXXREFERENCEPLANEX.XXXX mmREFERENCE PLANEX.XXXX mmDisplays the reference plane distance RDD?5: XXX.XXX

Strona 9

Menu Name Function Description GPIB MnemonicCHx-SXXREFERENCEPLANEX.XXXX mm 6REFERENCE PLANEX.XXXX mmDisplays the reference plane distance RDD?9: XXX.X

Strona 10 - Static Safety

Menu Name Function Description GPIB MnemonicFILTER SETUPBANDWIDTH LOSSVALUE XX.XXX dBA loss is a positive number. A value of 0 to999.999 dB may be ent

Strona 11 - Narrative Table of Contents

Menu Name Function Description GPIB MnemonicMARKERS(Power Sweep)READOUTMARKERSChanges the soft-key display to one of theREADOUT MARKER menu.NoneMARKER

Strona 12

Menu Name Function Description GPIB MnemonicMARKERREADOUTFUNCTIONS(Power Sweep)MARKERS ONACTIVE CHANNELSelects normal markers on active channel marker

Strona 13 - Table of Contents

Narrative Table of ContentsChapter 1 — General InformationThis chapter provides a general description of the Anritsu Model MS462XX Vector Network Meas

Strona 14 - Table of Contents (Continued)

Menu Name Function Description GPIB MnemonicSEARCHVALUE XX.XX dB Target search value. A value from -999.999 to999.999 dB may be entered.SRCH; SRCH?SET

Strona 15

Menu Name Function Description GPIB MnemonicSEARCH(X-AXIS)Y VALUEXX.XXX DBEnters the marker search Y-value. SRCH; SRCH?X VALUEXX.XXX DBMDisplays the m

Strona 16

4-8 FREQ KEY MENUS This key provides entry to frequency functions.Menu Name Function Description GPIB MnemonicDISCRETE FILLSTART FREQXXX.XXXXXXXX GHzE

Strona 17

Menu Name Function Description GPIB MnemonicFREQUENCY(In SegmentedSweep FrequencyMode)(MODELS WITHOPTION 24 ONLY)STARTXXX.XXXXXXXX MHZEnter the start

Strona 18

Menu Name Function Description GPIB MnemonicINDIVIDUALFREQ INSERTNEXT FREQUENCYXXX.XXXXXXXX GHzEnter the individual frequency, in GHz. DFQ; DFQ?INSERT

Strona 19 - General Information

4-9 SWEEP KEY MENUS This key provides entry to frequency sweep functions.Menu Name Function Description GPIB MnemonicAPPLYSEGMENTEDSWEEP(MODELS WITHOP

Strona 20

Menu Name Function Description GPIB MnemonicDEFINESEGMENTEDSWEEP 2(MODELS WITHOPTION 24 ONLY)SEGMENT ON/OFF1Toggles the define sweep segment on or off

Strona 21

Menu Name Function Description GPIB MnemonicSEGMENTEDSWEEPDISCRETE FILL(MODELS WITHOPTION 24 ONLY)START FREQXXX.XXXXXXXX MHZEnters the discrete fill s

Strona 22 - INTERFACES INFORMATION

Menu Name Function Description GPIB MnemonicSEGMENTEDSWEEP LOG(MODELS WITHOPTION 24 ONLY)PRINT LOG Sends the segmented sweep log to a printer. PSGLOGS

Strona 23 - INFORMATION COMPONENT KITS

Menu Name Function Description GPIB MnemonicTRIGGERSMEASUREMENT NoneINTERNAL Selects internal sweep trigger. TIN; TXX?EXTERNAL Selects external sweep

Strona 24 - 1-8 PRECISION COMPONENT

Chapter 11 — Calibration KitsThis chapter provides a description and listing of components for the calibration kits.Appendix A — Sequence OperationThi

Strona 25

4-10 POWER KEY MENUS This key provides entry to power functions.Menu Name Function Description GPIB MnemonicCLEAR ALLEXTD. RCVR(MODELS WITHOPTION 24)C

Strona 26 - 1-10 PERFORMANCE

Menu Name Function Description GPIB MnemonicFLAT TEST PORTPOWERCALIBRATIONFLAT POWERCALIBRATIONNoneCALIBRATING TESTPORTNonePRESS <CLEAR> TOABORT

Strona 27 - 1-11 PREVENTIVE

Menu Name Function Description GPIB MnemonicPOWER 2(Power Sweep)SELECT SOURCEXSelect source. PSRC; PSRC?CENTERXXX.XXXXXXXX dBmEnter the center power i

Strona 28 - INFORMATION

Menu Name Function Description GPIB MnemonicRECEIVERCALIBRATIONCONNECT ATHROUGH LINEBETWEEN TESTPORTSNoneRECEIVER TYPETEST/REFERENCE*Selects the recei

Strona 29 - Installation

Menu Name Function Description GPIB MnemonicRECEIVERSETUPRECEIVER CAL Changes the soft-key display to the RECEIVERCALIBRATION menu.NoneRECEIVER TYPETE

Strona 30

Menu Name Function Description GPIB MnemonicSOURCE 1SETUPSOURCE 1 POWERXX.XX dBm(5 TO -15 dB)Enter and display the power level, in dBm, of theSource 1

Strona 31

Menu Name Function Description GPIB MnemonicSOURCE 4SETUPSOURCE 4 POWERXX.XX dBmSets external Source 4 power level. PW4; PW4?RETURN Returns to the pre

Strona 32 - 2-4 GPIB SETUP AND

4-11 CONFIG KEY MENUS This key provides entry to configuration functions.Menu Name Function Description GPIB Mnemonic2 PORT DUTEMBEDDINGDE-EMBEDDINGSE

Strona 33 - 2-6 ETHERNET SETUP AND

Menu Name Function Description GPIB Mnemonic(ADDON/MODIFY)ADDON/MODIFY Informational text NoneEMBEDDING/DE-EMBEDDINGLC CKT TYPECIRCUITTOPOLOGY(ADDON/M

Strona 34 - 2-8 RACK MOUNT

Menu Name Function Description GPIB MnemonicDATA POINTS1601 MAX POINTS Selects measurement data points to be 1601. NP1601; FHI; ONP801 MAX POINTS Sele

Strona 35 - 2-9 PREPARATION FOR

Table of ContentsChapter 1 General Information1-1 SCOPE OF MANUAL ...1-31-2 INTRODUCTION ...

Strona 36 - SERVICE INSTALLATION

Menu Name Function Description GPIB MnemonicEMBEDDINGDE-EMBEDDINGLOG (PORT X)PRINT LOG Prints the embedding/de-embedding log. OEDELOGSAVE LOG TOFLOPPY

Strona 37 - INSTALLATION SERVICE

Menu Name Function Description GPIB MnemonicHOLDHOLD FUNCTIONHOLD/ Causes the hold key (button) to alternately stop andcontinue the sweep from the sto

Strona 38

Menu Name Function Description GPIB MnemonicLC CIRCUIT(ADDON/MODIFY)For 2-port device.SELECT LC CIRCUITTYPE (XXXX)Changes the soft-key display to the

Strona 39 - A Primer

Menu Name Function Description GPIB MnemonicNTWKEMBEDDINGDE-EMBEDDINGEMBEDDING/DE-EMBEDDINGON/OFFTurns EMBEDDING/DE-IMPEDDING On or Off. EDE0; EDE1; E

Strona 40

Menu Name Function Description GPIB MnemonicSAVE/RECALLSETUPSAVE SETUP ONHARD DISK Saves setup to hard drive. NoneFLOPPY DISK Saves setup to floppy dr

Strona 41

Menu Name Function Description GPIB MnemonicSOURCECONFIGURATIONSOURCE 1 Changes the soft-key display to the SOURCE 1CONFIGURATION menu.NoneSOURCE 2 Ch

Strona 42 - TERMINATION

Menu Name Function Description GPIB MnemonicT-LINE SECTION(ADDON/MODIFY)IMPEDANCEXXXXXXXXXXXXSets the impedance of the T-line section. EDEIMP; EDEIMP?

Strona 43 - A PRIMER ANALYZERS

4-15 START PRINT KEYMENUSThis key provides entry to printer, plotter, or disk file functions.Menu Name Function Description GPIB MnemonicSELECT FILE T

Strona 44 - ANALYZERS A PRIMER

4-16 UTILITY KEY MENUS This key provides entry to utility functions.Menu Name Function Description GPIB MnemonicAUTOCALASSURANCEPORT CONFIG(1X2A3B4C O

Strona 45 - MS462XX OM 3-7

Menu Name Function Description GPIB MnemonicAUTOCALASSURANCELIMITS 2Displays with a 2-port AutoCal box.LOAD MATCHDC - 2 GHZ37.0 DB2-6 GHZ37.0 DBChange

Strona 46 - 3-8 MS462XX OM

2-9 PREPARATION FOR STORAGE AND/OR SHIPMENT...2-7Preparation for Storage ...2-7Preparation for Shipment ...

Strona 47

Menu Name Function Description GPIB MnemonicAUTOCALASSURANCELIMITSDisplays in MS4623X with a 4-port AutoCal box.DIRECTIVITYDC - 3 GHZ40.0 DB3-6GHZ35.0

Strona 48

Menu Name Function Description GPIB MnemonicAUTOCALASSURANCELIMITSDisplays in MS4624X with a 4-port AutoCal box.DIRECTIVITYDC - 3 GHZ40.0 DB3-6GHZ35.0

Strona 49

Menu Name Function Description GPIB MnemonicAUTOCALASSURANCEMANUALCONTROLSELECT BOX TYPE2 PORT/4 PORTToggles between AutoCal box types, 2 or 4-port. A

Strona 50

Menu Name Function Description GPIB MnemonicAUTOCAL CHAR-ACTERIZATIONDisplayed for a model MS4622X with a 4-port AutoCal.(1X2C)CONNECT AUTOCALPORT C T

Strona 51 - Front Panel Operation

Menu Name Function Description GPIB MnemonicCLOCK SETUPYEAR XXXX Sets the YEAR. DATE; DATE?MONTH XX Sets the MONTH. DATE; DATE?DAY XX Sets the DAY. DA

Strona 52

Menu Name Function Description GPIB MnemonicCOMMAND LINEABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789()-!#$%’@^_’{}~*?:\.SPEnter a DOS command line using the r

Strona 53

Menu Name Function Description GPIB MnemonicFLOPPY DISKUTILITIESDISPLAY DIRECTORY Directory displays in the screen data area. DIRDELETE FILES Change t

Strona 54 - KEY DESCRIPTIONS OPERATION

Menu Name Function Description GPIB MnemonicH/WCALIBRATIONDISKOPERATIONSRECALL FROM HARDDISKChanges the soft-key display to the SELECT FILETO READ men

Strona 55 - OPERATION KEY DESCRIPTIONS

Menu Name Function Description GPIB MnemonicINSTRUMENTSTATEPARAMETERSSYSTEM Displays all of the system parameters. DGSCALIBRATION Displays the calibra

Strona 56 - 4-3 SOFT-KEY MENUS

Menu Name Function Description GPIB MnemonicPERIPHERALTESTSLCD DISPLAY Provides a graphic display for evaluating screencolors and linearity.DGT1; DGT2

Strona 57 - FRONT PANEL APPL

Chapter 5 Error and Status Messages5-1 INTRODUCTION ...5-35-2 ERROR MESSAGES...5-3Chapt

Strona 58 - APPL FRONT PANEL

Menu Name Function Description GPIB MnemonicREAR PANEL TTLOUTPUT MODESELECT PORT None1 Selects Port 1. RVTP1; RVTPX?2 Selects Port 2. RVTP2; RVTPX?3 S

Strona 59

Menu Name Function Description GPIB MnemonicSELECT FILE TODELETEFILE 1 Selects FILE 1 data to be deleted. NoneFILE 2 Selects FILE 2 data to be deleted

Strona 60

Menu Name Function Description GPIB MnemonicTROUBLESHOOTING 2(SERVICE USEONLY)HET OSCILLATORVOLTAGESelects HET OSCILLATOR VOLTAGE. TSHETOSOURCE 1 Chan

Strona 61

Menu Name Function Description GPIB MnemonicTROUBLE-SHOOTINGEXTENDEDFLASH MEMORY!!!WARNING!!!THIS TEST TAKESABOUT 4 MINUTES ANDDESTROYS THEINSTRUMENT’

Strona 62

Menu Name Function Description GPIB MnemonicTROUBLESHOOTINGSOURCE 1DDS REFERENCECLOCK VOLTAGESelects DDS REFERENCE CLOCK VOLTAGE forSOURCE 1.TSDDSS1OF

Strona 63

Menu Name Function Description GPIB MnemonicTROUBLESHOOTING SRAMDISK!!!WARNING!!!THIS TEST DESTROYSTHE HARMONIC CALCOEFFICIENTS THEHARD DISK ISCORRUPT

Strona 64

Menu Name Function Description GPIB MnemonicTYPE OF FILESTO COPYFRONT PANEL SETUPAND CAL DATAChanges the soft-key display to the SELECT FILETO COPY me

Strona 65

Menu Name Function Description GPIB MnemonicUTILITYINSTRUMENT STATEPARAMETERSChanges the soft-key display to the INSTRUMENTSTATE PARAMETERS menu.NoneR

Strona 66

4-17 HARD COPY KEYMENUSThis key provides entry to hard copy functions.Menu Name Function Description GPIB MnemonicDISK FILEOPTIONSDESTINATION NoneHARD

Strona 67

Menu Name Function Description GPIB MnemonicM3P SETUPSELECTBALANCEDPORT PAIR(MODELS WITHOPTION 24)(1:2)/(2:1) Selects Ports 1:2 or 2:1 as the balanced

Strona 68

7-4 MEASUREMENT CALIBRATION—SLIDING TERMINATION ...7-137-5 STANDARD 2-PORT (OSL) CALIBRATION PROCEDURE ...7-197-6 TRANSMISSION AND REFL

Strona 69

Menu Name Function Description GPIB MnemonicM4P SETUPSELECT 2NDPAIR(1:2)/(2:1) Selects Ports 1:2 or 2:1 as the balanced port pair forthe M4P setup.M4P

Strona 70 - MEAS FRONT PANEL

Menu Name Function Description GPIB MnemonicOUTPUTHEADERS SETUP3DATE ON/OFF Toggles date on or off. LDT1; LDT0SEQ HELP MESSAGEON/OFFToggles the Sequen

Strona 71 - FRONT PANEL MEAS

Menu Name Function Description GPIB MnemonicPLOT SETUPFULL PLOT The plotter will plot everything displayed on thescreen (data traces, graticule, menu

Strona 72

Menu Name Function Description GPIB MnemonicPRINTER SETUPGRAPHICALDATAHEADER ON/OFF Toggles printing of header on or off. HD1; HD0; HDX?SCREEN AREA TO

Strona 73

Menu Name Function Description GPIB MnemonicSNP OPTIONSPORT 1 Select S1P format for Port 1. OS1P1PORT 2 Select S1P format for Port 2. OS1P2PORT 3 Sele

Strona 74

4-18 DEFAULT KEY MENUS This key provides entry to default functions.Menu Name Function Description GPIB MnemonicDEFAULTCONTINUE Erases current instrum

Strona 75

4-19 SAVE/RECALL KEYMENUSThis key provides entry to save and recall functions.Menu Name Function Description GPIB MnemonicRECALLFRONT PANEL SETUPIN IN

Strona 76

Menu Name Function Description GPIB MnemonicRECALL FRONTPANEL SETUPFROM INTERNALMEMORYMEMORY 1 * Causes the current front panel setup to be recalledfr

Strona 77

Menu Name Function Description GPIB MnemonicSAVELINEAR CAL NoneON HARD DISK Changes the soft-key display to the SELECT FILETO READ menu.SAVEON FLOPPY

Strona 78 - DISPLAY FRONT PANEL

Menu Name Function Description GPIB MnemonicSAVE FRONTPANEL SETUP TOINTERNALMEMORYMEMORY 7 * Same as above, except the setup saves to memorylocation 7

Strona 79 - FRONT PANEL DISPLAY

10-5 MS462XA/B NON-RATIO PARAMETERS TEST...10-510-6 MS462XA/B HIGH LEVEL NOISE TEST ...10-1010-7 MS462XC SYSTEM DYNAM

Strona 80

Menu Name Function Description GPIB MnemonicSELECT FILE-NAMEABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789()-!#$%’@^_’{}~*?:\.SPName your filename using the rot

Strona 81

4-12 SEQ KEY MENUS This key provides entry to sequence functions.Menu Name Function Description GPIB MnemonicCONDITIONSTATEMENTIF LIMIT TEST PASS Chan

Strona 82

Menu Name Function Description GPIB MnemonicDELETESEQUENCESEQUENCE 1XXXXXRemove sequence 1 from SRAM. SEQDEL1SEQUENCE 2XXXXXRemoves sequence 2 from SR

Strona 83

Menu Name Function Description GPIB MnemonicENTER ASEQUENCEOPERATORMESSAGEor(ENTER ASEQUENCECOMMENT)(Message or comment) Display the message text. Non

Strona 84

Menu Name Function Description GPIB MnemonicGOSUBSEQUENCESEQUENCE 1XXXXXSelects sequence 1 to be called during sequenceexecution; the corresponding st

Strona 85

Menu Name Function Description GPIB MnemonicRENAMESEQUENCESEQUENCE 1XXXXXChanges the soft-key display to the RENAME ASEQUENCE menu.SEQNAM1; SEQNAM1?SE

Strona 86

Menu Name Function Description GPIB MnemonicSAVE SEQUENCESEQUENCE 1XXXXXSelects sequence 1 to save on hard/floppy disk; astatement is added to the seq

Strona 87

Menu Name Function Description GPIB MnemonicSELECT FILE TOREADFILE 1 SEQ Displays the data stored in file number 1. NoneFILE 2 SEQ Displays the data s

Strona 88

Menu Name Function Description GPIB MnemonicSEQUENCENAME(Name) Display the name. NoneABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789()-!#$%&’@^_’{}-*?:\.SPNa

Strona 89

4-13 CAL KEY MENUS This key provides entry to calibration functions.Menu Name Function Description GPIB Mnemonic2 PORTCALIBRATION(MODELS WITHOPTION 24

Strona 90

A-18 TTL OUTPUT FOR CONTROLLING PERIPHERALS ...A-18A-19 TTL INPUT CONTROL STATEMENT ...A-19A-20 BNC TTL OUTPUT CONTROL

Strona 91

Menu Name Function Description GPIB Mnemonic3 PORTCALIBRATIONCONNECTCALIBRATION DEVICENoneSLIDE LOAD TOPOSITION xSlide the load to the indicated posit

Strona 92

Menu Name Function Description GPIB Mnemonic4 PORTCALIBRATION(SOLT/SOLR)CONNECT ARECIPROCALBETWEEN PORTX AND YWhen a 3 Port calibration is already com

Strona 93

Menu Name Function Description GPIB Mnemonic4 PORTCALIBRATIONCONNECTCALIBRATION DEVICENonePORT 3: OPENPORT 4: SHORTConnect specified open to Port 3 an

Strona 94

Menu Name Function Description GPIB Mnemonic4 PORTCONFIRMCALIBRATIONPARAMETERS(SOLT/SOLR)PORT 3 CONNXXXXXXXXXXDisplays the current connector type and

Strona 95

Menu Name Function Description GPIB Mnemonic4 PORTTHROUGHLINEOFFSET LENGTH(TRX)PORT 1, 3XXX mmEnter Port 1, 3 THROUGH offset length for 4 portcalibrat

Strona 96

Menu Name Function Description GPIB Mnemonic4 PORTSELECT TYPE OFLOADBROADBANDFIXED LOADSelects a broadband load for calibration and changessoft-key di

Strona 97

Menu Name Function Description GPIB MnemonicADAPTERREMOVAL12-TERM CAL X AND YMUST EXIST IN THECURRENT DIRECTORYNoneELECTRICAL LENGTHOF THE ADAPTERXXX.

Strona 98

Menu Name Function Description GPIB MnemonicAUTOCAL 4SELECT CALIBRATIONTYPENoneFULL 2 PORT Selects a traditional 2-portShort-Open-Load-Thru(SOLT) cali

Strona 99

Menu Name Function Description GPIB MnemonicAUTOCAL PORTCONFIGURATION(FULL 2 PORT CAL/2 PORT REFLEC-TION/ADAPTERREMOVAL)PORT CONFIG.1L2R/1R2LToggles b

Strona 100 - KEY MENUS OPERATION

Menu Name Function Description GPIB MnemonicAUTOCAL PORTCONFIGURATION(FULL 3 PORT CAL)SELECT VNA PORTFOR AUTOCAL PORT X1/2/3This menu displays for a 4

Strona 101 - OPERATION KEY MENUS

Table of Contents1-1 SCOPE OF MANUAL ...1-31-2 INTRODUCTION ...1-31-3 SERIAL NUMBER...

Strona 102

Menu Name Function Description GPIB MnemonicAUTOCAL PORTCONFIGURATION(FULL 4 PORT CAL)SELECT VNA PORTFOR AUTOCAL PORT X1/2/3/4This menu displays for a

Strona 103

Menu Name Function Description GPIB MnemonicAUTOCAL SETUP(FULL 4 PORT CAL)PORT CONFIG.(1X2A3B4C)Displays only if connected to a 4-port AutoCal box.Cha

Strona 104

Menu Name Function Description GPIB MnemonicAUTOCAL SETUP(ADAPTERREMOVAL)PORT CONFIG.(1X2A)Displays if selected box type is a 4-port AutoCal.Changes t

Strona 105

Menu Name Function Description GPIB MnemonicBEGIN 2 PORTCALIBRATION(12-TERMS)KEEP EXISTING CALDATAChanges the soft-key display to the FREQUENCYmenu.KE

Strona 106

Menu Name Function Description GPIB MnemonicCAL FREQ RANGESTARTXXX.XXXXXXXXXGHzEnter the sweep-start frequency for calibration. Ifdesired; frequency c

Strona 107

Menu Name Function Description GPIB MnemonicCALIBRATIONCOMPONENTUTILITIESINSTALL KIT INFOFROM FLOPPY DISKInstalls the information from the calibration

Strona 108

Menu Name Function Description GPIB MnemonicCALIBRATIONSEQUENCE2CONNECT THERECIPROCALXXXXXX mmBETWEENTEST PORTSNoneMEASURE FORWARD((NO) CAL EXISTS)Mea

Strona 109

Menu Name Function Description GPIB MnemonicCHANGE CALMETHOD AND LINETYPECAL METHOD NoneSTANDARD(NOT USED FORWAVEGUIDE)or with Option 24:SOLT/SOLR(NOT

Strona 110

Menu Name Function Description GPIB MnemonicCHANGE LRL/LRMPARAMETERS2CHARACTERIZECAL DEVICESNoneDEVICE 1LINE 1 (REF)X.XXXX mmEnter length of line 1. L

Strona 111

Menu Name Function Description GPIB MnemonicCHANGEPARAMETERS2CHARACTERIZECAL DEVICESNoneDEVICE 1 P3-P1LINE LENGTHX.XX mmSets device 1 line length. LL1

Strona 113 - FRONT PANEL FREQ

Figure 1-1. Model MS462XX Vector Network Measurement System

Strona 114

Menu Name Function Description GPIB MnemonicCONFIRMCALIBRATIONPARAMETERS2(STANDARDMICROSTRIP)PORT 1OPEN/SHORTChanges the soft-key display to the PORT

Strona 115 - FRONT PANEL SWEEP

Menu Name Function Description GPIB MnemonicCONFIRMCALIBRATIONPARAMETERS4(OPEN SHORTMICROSTRIP)OFFSET LENGTHSOF SHORTSChanges the soft-key display to

Strona 116 - SWEEP FRONT PANEL

Menu Name Function Description GPIB MnemonicCONFIRMCALIBRATIONPARAMETERS7(LRL/LRMMICROSTRIP)LRL/LRM PARAMETERS Changes the soft-key display to the CHA

Strona 117

Menu Name Function Description GPIB MnemonicCUSTOMIZE CAL(FULL 2 PORT CAL)SET ALL ON Toggles all S-parameter selections on. FXSONSET ALL OFF Toggles a

Strona 118

Menu Name Function Description GPIB MnemonicCUSTOMIZE CAL 2(FULL 4 PORT CAL)S13, TRANS ON/OFF Toggles the S13 selection on or off. FXS13T0;FXS13T1; FX

Strona 119

Menu Name Function Description GPIB MnemonicDATA POINTS1601 MAX POINTS Provides 1601 data points across selected frequencyrange.FHI; NP1601; ONP801 MA

Strona 120 - POWER FRONT PANEL

Menu Name Function Description GPIB MnemonicDISPLAYINSTALLED TESTPORT CONNECTORINFO 2GPC - 3.5 (M) Displays coefficient data for the GPC-3.5 malecompo

Strona 121 - FRONT PANEL POWER

Menu Name Function Description GPIB MnemonicDISPLAYINSTALLED TESTPORT CONNECTORINFO 5SPECIAL (M) Displays coefficient data for the SPECIAL connectorma

Strona 122

Menu Name Function Description GPIB MnemonicENTER THETHROUGH LINEPARAMETERSOFFSET LENGTHXX.XX mmEnter offset length of throughline device. TOL; TOL?TH

Strona 123

Menu Name Function Description GPIB MnemonicFULL TERM CAL(FULL -PORT CAL)PORT 1 ONPORT 2 ONDisplayed when a 12-term cal exists.Toggles port 1 selectio

Strona 124

Chapter 1General Information1-1 SCOPE OF MANUAL This manual provides general information, installation, and operatinginformation for the MS4622, MS462

Strona 125

Menu Name Function Description GPIB MnemonicGAINCOMPRESSIONCALIBRATIONCALIBRATE FORFLATNESS((NO) CAL EXISTS)Changes the soft-key display to the SOURCE

Strona 126

Menu Name Function Description GPIB MnemonicHARMONICENHANCEMENTCALIBRATIONCONNECT ATHROUGHLINEBETWEEN TESTPORTSNoneABORT CAL Changes the soft-key disp

Strona 127 - FRONT PANEL CONFIG

Menu Name Function Description GPIB MnemonicHARMONIC PHASECALIBRATIONCONNECT STANDARDEXTERNAL DEVICEBETWEEN TESTPORTSNoneABORT CAL Changes the soft-ke

Strona 128 - CONFIG FRONT PANEL

Menu Name Function Description GPIB MnemonicIMDCALIBRATION 2CONNECT ATHROUGHLINEBETWEEN SOURCEPORT AND RECEIVERPORT (PORT 2)NoneABORT CAL Aborts the c

Strona 129

Menu Name Function Description GPIB MnemonicNETWORKEXTRACTIONTHE ‘OUTER’ AND‘INNER’ CALIBRATIONFILES MUST EXIST INTHE CURRENTDIRECTORYGENERATE S2P FIL

Strona 130

Menu Name Function Description GPIB MnemonicNOISE FIGURECALIBRATION3RESTARTMEASUREMENTChanges the soft-key display to the NOISE FIGURECALIBRATION 2 me

Strona 131

Menu Name Function Description GPIB MnemonicPORT XOPEN DEVICEENTER THECAPACITANCECOEFFICIENTSNoneTERM 1-C0P/MXXX.XXe-15Enter the term 1 coefficient va

Strona 132

Menu Name Function Description GPIB MnemonicRECEIVERCALIBRATION(SWEPT POWER/FREQUENCY GAINCOMPRESSIONAPPLICATIONS)CONNECT A THROUGHLINE BETWEEN TESTPO

Strona 133

Menu Name Function Description GPIB MnemonicSELECT1 PATH 2 PORTCALIBRATIONTYPEFORWARD PATH(S11, S21)Corrects the forward transmission and reflection e

Strona 134

Menu Name Function Description GPIB MnemonicSELECT DEVICETYPE(4 PORT CAL WITHEXISTING 3 PORTCAL)1,4 DEVICE TYPETHRU/RECIPROCALSelect the device type f

Strona 135

1-6 SYSTEM DESCRIPTION The MS462XX Vector Network Measurement System (VNMS)(Figure 1-1) is a single-instrument system that contains a built-insource,

Strona 136

Menu Name Function Description GPIB MnemonicSELECT DEVICETYPE(4 PORT CALWITHOUT EXISTING3 PORT CAL)1,3 DEVICE TYPETHRU/RECIPROCALSelect the device typ

Strona 137 - 4-15 START PRINT KEY

Menu Name Function Description GPIB MnemonicSELECT DEVICETYPE(4 PORT CALWITHOUT EXISTING3 PORT CAL)2,4 DEVICE TYPETHRU/RECIPROCALSelect the device typ

Strona 138 - UTILITY FRONT PANEL

Menu Name Function Description GPIB MnemonicSELECT PORT XCONNECTORTYPE3TNC (M) Sets port x to be male TNC connector. CMC; P1C?; P2C?TNC (F) Sets port

Strona 139 - FRONT PANEL UTILITY

Menu Name Function Description GPIB MnemonicSELECTTRANSMISSIONFREQ RESPONSECALIBRATIONTYPEFORWARD PATH (S21) Changes the soft-key display to the SELEC

Strona 140

Menu Name Function Description GPIB MnemonicSOURCE 1 FLATTEST PORTPOWERCALIBRATIONXXXX POINTSMEASURE 1 PWRPOINT EVERY XXPOINT(S)Displays the number of

Strona 141

Menu Name Function Description GPIB MnemonicTHRUASSIGNMENT(FULL 3 PORT CAL)THRU (1-2)*AUTOCAL/THRULINE LENGTH0.000 mmSelect thru type for port 1-2. En

Strona 142

Menu Name Function Description GPIB MnemonicTHRUASSIGNMENT2(FULL 4 PORT CAL)THRU (2-3)*OMIT/THRULINE LENGTH0.000 mmSelect whether to omit the thru or

Strona 143

Menu Name Function Description GPIB MnemonicTHRU UPDATE(FULL 3 PORT CAL)THRU (1-2)LINE LENGTH0.000 mmSet the line length offset for the true thru betw

Strona 144

Menu Name Function Description GPIB MnemonicTHRU UPDATE 2(FULL 4 PORT CAL)THRU (2-3)OMIT/THRULINE LENGTH0.000 mmSelect whether to omit the thru or sel

Strona 145

Menu Name Function Description GPIB MnemonicTHRU/RECIPROCALPARAMETERS(3 PORTCALIBRATIONWITH OPTION 24)MEASURE PORT 1,3?OMIT/INCLUDESelect whether to i

Strona 146

NOTESThe MS462XC VNMS- Direct Receiver Access versioncannot make system measurements without an externaltest set that couples power into the reference

Strona 147

Menu Name Function Description GPIB MnemonicTHRU/RECIPROCALPARAMETERS(4 PORT CAL WITH3 PORT CAL)MEASURE PORT 1,3?OMIT/INCLUDESelect whether to include

Strona 148

Menu Name Function Description GPIB MnemonicTRANSMISSIONAND REFLECTIONCALIBRATIONDisplayed for MS4624X modelsAPPLY CAL2-PORT(CAL EXISTS)Selects a 2-po

Strona 149

4-14 AVG KEY MENUS This key provides entry to averaging and IF bandwidth functions.Menu Name Function Description GPIB MnemonicAVERAGE 1AVERAGE ON/OFF

Strona 150

Menu Name Function Description GPIB MnemonicAVERAGE 3Displays in the Transmission/Reflection Calibration Mode Only.AVERAGE ON/OFF Toggles the averagin

Strona 151

4-184 MS462XX OMFUNCTION KEYS FRONT PANELTEMPLATE KEY DESCRIPTIONSCtrlAltShiftCtrlAltShiftCh 1 Ch 2 Ch 3 Ch 4Freq SweepPower ConfigSoftkey 1 Softkey 2

Strona 152

Table of Contents5-1 INTRODUCTION ...5-35-2 ERROR MESSAGES...5-3Chapter 5Error and Stat

Strona 154

Chapter 5Error and StatusMessages5-1 INTRODUCTION This chapter lists, describes, and provides corrective action for theerror messages that appear on t

Strona 155

ERROR ERROR AND STATUSMESSAGES MESSAGES5-4 MS462XX OMCode Message Description Corrective ActionLANC BCMP ERROR LANC controller test fail Contact facto

Strona 156

ERROR AND STATUS ERRORMESSAGES MESSAGESMS462XX OM 5-5Code Message Description Corrective ActionSCSI FIFO DMA PARITY BIT ERROR Hard drive controller te

Strona 157

1-8 PRECISION COMPONENTKITSTwo types of precision-component kits are available: calibration andverification. Calibration kits contain components used

Strona 158 - 4-17 HARD COPY KEY

ERROR ERROR AND STATUSMESSAGES MESSAGES5-6 MS462XX OMCode Message Description Corrective ActionSKT INVALID DESCRIPTOR Socket has been removedSKT INCOM

Strona 159 - FRONT PANEL HARD COPY

ERROR AND STATUS ERRORMESSAGES MESSAGESMS462XX OM 5-7Code Message Description Corrective ActionTOO HIGH TEMP CHANGE Temperature is too high tochange t

Strona 160 - HARD COPY FRONT PANEL

ERROR ERROR AND STATUSMESSAGES MESSAGES5-8 MS462XX OMCode Message Description Corrective ActionPOWER RESTORED TO CAL RANGE Power restored is within ca

Strona 161

ERROR AND STATUS ERRORMESSAGES MESSAGESMS462XX OM 5-9Code Message Description Corrective ActionSource Error Codes6129 EXT SRC ERROR6130 EXT SRC LIST E

Strona 162

ERROR ERROR AND STATUSMESSAGES MESSAGES5-10 MS462XX OMCode Message Description Corrective Action7100 FILE MARKED READ ONLY Can't perform a write

Strona 163

ERROR AND STATUS ERRORMESSAGES MESSAGESMS462XX OM 5-11Code Message Description Corrective Action7350 EXT TRIG RATE TOO FAST Pulse rate of the external

Strona 164

ERROR ERROR AND STATUSMESSAGES MESSAGES5-12 MS462XX OMCode Message0002 SRC1 COMM FAIL0003 SRC2 COMM FAIL0005 PWR SUPPLY UP +5V0006 PWR SUPPLY UP +15V0

Strona 165 - FRONT PANEL DEFAULT

ERROR AND STATUS ERRORMESSAGES MESSAGESMS462XX OM 5-13/5-14Code Message0327 SRC2 FREQ PATH BAND8 FAIL0328 SRC2 FREQ PATH BAND9 FAIL0329 SRC2 FREQ PATH

Strona 167 - FRONT PANEL SAVE/RECALL

Table of Contents6-1 INTRODUCTION ...6-36-2 DISPLAY MODES AND TYPES ...6-3Single Channel Displ

Strona 168 - SAVE/RECALL FRONT PANEL

3753RCalibration KitThe 3753R Calibration Kit (Figure 1-4) containsprecision Type N components with characteristicsthat are traceable to the NIST. Use

Strona 170

Chapter 6Data Displays6-1 INTRODUCTION This chapter provides discussion and examples of the various types ofdata displays.6-2 DISPLAY MODES ANDTYPESTh

Strona 171 - FRONT PANEL SEQ

Dual ChannelDisplay—Ch1and3orCh 2 and 4If you have chosen a dual display of magnitude andphase, the affected area of the LCD screen is subdi-vided int

Strona 172 - SEQ FRONT PANEL

Four ChannelDisplay—Ch 1, 2, 3, 4From four-to-eight graph types are displayed. Ineach quadrant, the graph type can be any of the pos-sible choices lis

Strona 173

Dual TraceOverlayFor rectilinear graph types, two traces can be dis-played, one overlaid (superimposed) on the other(Figure 6-5). By soft-key selectio

Strona 174

Four ChannelOverlayFor rectilinear graph types, four traces can be dis-played, each one overlaid (superimposed) on another(Figure 6-6). Each trace is

Strona 175

Graph DataTypesThe data types (real, imaginary, magnitude, phase)used in the displayed graph-types reflect the possi-ble ways in which S-Parameter dat

Strona 176

DATA DISPLAY MODESDISPLAYS AND TYPESMS462XX OM 6-9GRAPH TYPELINEAR POLARSMITH CHART(IMPEDANCE)SMITH CHART(ADMITTANCE)GROUP DELAYPOWER OUTSWRMORERETURN

Strona 177

DISPLAY MODES DATAAND TYPES DISPLAYS6-10 MS462XX OMGRAPH TYPELOG MAGNITUDELINEAR MAGNITUDEPHASEREALIMAGINARYLOG POLARMORERETURN90S21 REFLECTIONL0G POL

Strona 178

DATA DISPLAY MODESDISPLAYS AND TYPESMS462XX OM 6-11Figure 6-11. 3 dB Compressed Smith ChartFigure 6-12. 20 dB Expanded Smith Chart

Strona 179 - FRONT PANEL CAL

1-9 OPTIONS The following options are available.qOption 1: Rack Mount KitqOption 1A: Rack Mount Kit for use with System ConsoleqOption 2: Time (Distan

Strona 180 - CAL FRONT PANEL

6-3 FREQUENCY MARKERS Frequency markers are described below.MarkerAnnotationFigure 6-13 shows how the MS462XX annotatesmarkers for the different graph

Strona 181

above (below) the marker. The delta symbol appearsabove (below) the number.MarkerStatisticsThe marker statistics available in the marker menucalculate

Strona 182

Each channel can produce segmented limits. They allow differentupper and lower limit values to be set at up to ten segments across themeasurement rang

Strona 183

ScaleResolutionEach measurement display is annotated with thescale resolution. For log-magnitude displays resolu-tion ranges from 0.001 to 50 dB per d

Strona 184

6-6 DATA DISPLAYCONTROLData display control is described below.ActiveChannelSelectionFigure 6-16 shows the algorithm that the MS462XXuses to display t

Strona 185

Data DisplayUpdateWhen you change a control panel parameter that af-fects the appearance of the display, the entire dis-play changes immediately to re

Strona 186

change reflects the changes in measurement data atthe marker’s new frequency, using data stored fromthe previous sweep.Display ofMarkersOnce you have

Strona 187

Plotter Output The protocol used to control plotters is HP-GL(Hewlett-Packard Graphics Language). HP-GL con-tains a comprehensive set of vector graphi

Strona 188

HARD COPY DATAAND DISK OUTPUT DISPLAYS6-20 MS462XX OMFigure 6-17. Example of a Plot (MS4622B)

Strona 189

DATA HARD COPYDISPLAYS AND DISK OUTPUTMS462XX OM 6-21/6-22Figure 6-18. Example of a Plot (MS4623A)

Strona 190

1-11 PREVENTIVEMAINTENANCEThe MS462XX VNMS requires periodic cleaning of the cooling fanscreen to prevent an obstruction that could facilitate overhea

Strona 192

Table of Contents7-1 INTRODUCTION ...7-37-2 CONVENTIONS ...7-37-3 MEASUREMENT CALIBRA

Strona 193

NOTEThe MS462XC VNMS- Direct Receiver Access version cannot make systemmeasurements without an external test set that couples power into thereference

Strona 194

Chapter 7Measurement Calibration7-1 INTRODUCTION This section provides discussion and examples for performing ameasurement calibration. It also provid

Strona 195

The use of cables and/or adapters does not effect thefinal measurement result, if they were in place forthe calibration process. The vector error corr

Strona 196

This approach offers two advantages:qIt minimizes wear on the more expensivetest set and cable connectorsqIt provides a simple solution to measuringno

Strona 197

UnderstandingtheCalibrationSystemMeasurement errors must be reduced by a processthat uses calibration standards. The standards mostcommonly used are O

Strona 198

Error modeling and flowgraphs are techniques usedto analyze the errors in a system. Error models de-scribe the errors, while flowgraphs show how these

Strona 199

MEASUREMENT CALIBRATION MEASUREMENTDISCUSSION CALIBRATION7-8 MS462XX OMPORT 1 PORT 2RF INPORT 1 PORT 2PORT 1 PORT 1PORT 2PORT 2PORT 1 PORT 2PORT 1 POR

Strona 200

Measurement calibration using the MS462XX isstraightforward and menu directed. A short timespent in preparation and preplanning will make theprocess s

Strona 201

1-12 CONVENTIONS Throughout this manual, path names may be used to represent thekeystrokes for a desired action or procedure. The pathname beginswith

Strona 202

Ideal Z0terminations must consist of two parts, aperfect connector and an infinite-length perfecttransmission line that absorbs all of the RF energyth

Strona 203

Evaluating theCalibrationThe MS462XX provides an accurate representationof complex data. However, it can only provide accu-racy to the extent of the s

Strona 204

The verification of the kit components is straightforward. The components are first measured withthe MS462XX, then they are compared with thedata reco

Strona 205

7-4 MEASUREMENTCALIBRATION—SLIDINGTERMINATIONSliding terminations (loads) are the traditional Z0calibration-reference devices for Vector Network Measu

Strona 206

Step 2. Push the outer locking ring towards the gauge to ex-pose the center pin.Step 3.Take the 01-210 Ref Flat from the kit.Step 4.While holding the

Strona 207

Step 5. While pressing the Ref Flat against the center pin,check that the pointer aligns with the “0” mark. If itdoes not, loosen the bezel lockscrew

Strona 208

Step 9. Remove the flush short by holding its body and un-screwing its connector.Step 10.Install the gauge onto the end of the slidingtermination.Step

Strona 209

Step 13. Loosen the FINE LOCK ring and turn the FINEADJ ring to position the gauge pointer 2-3 small di-visions on the “–” side of zero.Step 14.Turn t

Strona 210

Step 17. Remove the 01-223 Female Adapter (“F ADAPTERFOR PIN GAUGE”) from the kit.Step 18.Install the female adapter over the center pin andscrew it i

Strona 211

7-5 STANDARD 2-PORT(OSL) CALIBRATIONPROCEDUREThe standard 2-port calibration for the MS462XX Vector NetworkMeasurement System uses an Open, a Short, a

Strona 212

Table of Contents2-1 INTRODUCTION ...2-32-2 INITIAL INSPECTION ...2-32-3 PREPARATION FOR

Strona 213

tion frequency range. C.W.: Data is taken at onepoint. N-DISCRETE FREQUENCIES (2 TO 1601POINTS): This selection lets you specify a discretenumber of f

Strona 214

Step 12.Press LOAD TYPE.The next soft-key set lets you select either of two loadtypes, broadband or sliding. Broadband loads areadequate for all but t

Strona 215

The next soft-key set lets you define the power level ofthe signals at the two test ports.(If you needed to calibrate the test port for powerflatness,

Strona 216

7-6 TRANSMISSION ANDREFLECTION 3-PORTCALIBRATIONOnce the two port calibration has been successfully performed, theuser may select one of presently two

Strona 217

Step 3. Press the Cal (top left) then the following soft keys insequence:PERFORM 3 PORT CALCHANGE CAL METHODTRXNEXT CAL STEPNEXT CAL STEPCHANGE PARAME

Strona 218

Both three port completion steps will start with a two port calibration(first 12 terms) and will use the same isolation measurement steps toget the la

Strona 219

7-7 FOUR-PORTCALIBRATIONCalibration using the Four-Port Calibration technique is available for4-Port D models only. The following procedure calibrates

Strona 220

7-8 LRL/LRM CALIBRATIONPROCEDUREThe LRL/LRM (line-reflect-line/line-reflect-match) calibration* featureprovides an enhanced capability for error compe

Strona 221

When testing microstrip devices it is necessary to launch from coax tomicrostrip. In production testing this launching must be temporary, sothat the d

Strona 222

This soft-key set lets you select the number of fre-quency points at which calibration data is to betaken. The choices: NORMAL: Data is taken at up to

Strona 223

MS462XXVECTOR NETWORK MEASUREMENT SYSTEMOPERATION MANUALP/N: 10410-00203REVISION: NPRINTED: SEPTEMBER 2005COPYRIGHT 2005 ANRITSU CO.490 JARVIS DRIVEl

Strona 225

b.Press DEVICE 2 LINE/MATCH.Here you have another decision to make: whetheryour calibration is to be LRL or LRM. The soft keyacts as a toggle.c.If you

Strona 226

This indicates that your match is a high-band load.This load must have a passband such that it passesall frequencies from the breakpoint to the stop f

Strona 227

(If you needed to calibrate the test port for powerflatness, you would press the FLATNESS CORREC-TION soft key to ON.)Step 22.Press RESUME CAL.Step 23

Strona 228

LRL/LRM Calibration (Coaxial)An LRL cal kit is necessary to perform the coaxial calibration.Calibration kits for GPC-7 are available from Maury Microw

Strona 229

larger absolute difference, are used for the interval [f1, f2]. Data fromthe two lines, which make up the smaller absolute difference, are usedfor the

Strona 230

In each case, considerable time is lost due to reloading a 2-portcalibration between measurements or due to measuring 16 parameterseach time when only

Strona 231

Example Using Flexible CalibrationFigure 7-8 shows a 3-port device consisting of a 1-portresonator/antenna and a 2-port sub-DUT. If leakage is not of

Strona 232 - AVG FRONT PANEL

The light trace is the full 3-port calibration result while the dark traceuses the Flexible Calibration method. Aside from some minordifferences due t

Strona 234 - Function Keys

Table of Contents8-1 INTRODUCTION ...8-38-2 TIME DOMAIN MEASUREMENTS ...8-38-3 OPERATING TIME DO

Strona 235 - Messages

Chapter 2Installation2-1 INTRODUCTION This chapter provides information for the initial inspection andpreparation for use of the MS462XX Vector Networ

Strona 237

Chapter 8Time Domain8-1 INTRODUCTION This chapter describes the optional Time Domain feature.8-2 TIME DOMAINMEASUREMENTSThe Option 2, Time Domain feat

Strona 238 - MESSAGES MESSAGES

The lowpass impulse response displays the location of discontinuitiesas well as information useful in determining the impedance (R, L, orC) of each di

Strona 239 - ERROR AND STATUS ERROR

(TDR) you may feel more comfortable with step response, as thedisplays are similar.The lowpass-step response for a resistive impedance is a positive l

Strona 240

Use the bandpass-impulse response to show the location of adiscontinuity in time or distance, as indicated by changes in itsmagnitude. Unlike the lowp

Strona 241

Next, let us look at a complex circuit. A resistive impedance changeR<Z0and a shunt capacitance and series inductance. These impedancechanges are s

Strona 242

8-3 OPERATING TIMEDOMAINTo operate in the time domain mode, press the Appl key (below), thenthe TRANSMISSION AND REFLECTION and DOMAIN soft keys..The

Strona 243

Select SETUP then RANGE SETUP AND and use the START/STOPselections to set the range (Figure 8-12). Alternatively, you can useGATE SETUP and SPAN.TIME

Strona 244

For the lowpass mode select either IMPULSE or STEP Response andset the DC term. The MS462XXdefaults to the IMPULSE Responseand the AUTO EXTRAPOLATE mo

Strona 245

The Marker Range menu allows us to zoom in and display the rangebetween two selected markers (Figure 8-14).TIME OPERATINGDOMAIN TIME DOMAINMS462XX OM

Strona 246

2-4 GPIB SETUP ANDINTERCONNECTIONAll functions of the MS462XX (except power on/off and initialization ofthe hard disk) can be controlled remotely by a

Strona 247

8-4 WINDOWING Windowing is a frequency filter that we apply to the frequency-domaindata when we convert it to time-domain data. This filtering rolls o

Strona 248

8-5 GATING Gating is a time filter that allows for removing unwanted time-domainresponses by gating the desired response. We can view the isolatedresp

Strona 249 - Data Displays

8-6 ANTI-GATING Anti-gating is the opposite of gating. Whereby, gating provides forremoving all but the desired response, anti-gating displays all but

Strona 250

Table of Contents9-1 INTRODUCTION ...9-39-2 DESCRIPTION ...9-39-3 CALIBRATIONS...

Strona 251 - 6-2 DISPLAY MODES AND

AUTOCAL MODULEPowerOperateFigure 9-1. AutoCal Module, Power Supply, and CablesPower SupplyAutoCal ModuleCalibration CoefficientsDisk

Strona 252 - AND TYPES DISPLAYS

Chapter 9AutoCal9-1 INTRODUCTIONThis chapter provides a general description of the AutoCalcalibrators, including specifications, setup, and the use of

Strona 253 - DISPLAYS AND TYPES

Test Port Cable Converters (Anritsu series 36583) are used during andafter the calibration process to establish the desired test portconnector type.9-

Strona 254

isolation step requires a manual operation for the AutoCal module. Inorder to achieve high accuracy for the characterization of the leakage,a high ave

Strona 255

9-5 PHYSICAL SETUP See Figure 9-2 for an illustration of the connections necessary toperform an automatic calibration using the AutoCal module.There i

Strona 256

9-6 CHARACTERIZATIONFILESBefore performing an AutoCal on a MS462XX, the CharacterizationFile for the AutoCal Module must be loaded.Insert the AutoCal

Strona 257

2-5 SYSTEM GPIBINTERCONNECTIONThere are two rear panel GPIB IEEE-488 connectors. The IEEE 488.2connector is used to interface the MS462XX to an extern

Strona 258

menus again.CHANGE CAL SETUP lets you set up a new cali-bration, which is what we will do for this example.Step 9.Press the CHANGE CAL SETUP then FULL

Strona 259

Step 13.Ensure the AutoCal module is properly connectedbetween Ports 1 and 2, then press the START CALsoft key (AutoCal main menu).Follow the instruct

Strona 260 - MARKERS DISPLAYS

9-8 4-PORT AUTOCALThis section describes how to setup and use the 4-Port AutoCalmodule. Once setup, the use of the module is illustrated throughdiscus

Strona 261 - DISPLAYS

NOTEMS462XX firmware version 1.15 or later isrequired for automatic operation.Assurance Assurance is the process of validating a calibration.This is d

Strona 262 - CONTROLS DISPLAYS

frequency ranges will have a lower point densitythan may be desired. The resulting interpolation er-ror is typically very small but if an issue, you c

Strona 263 - DISPLAYS CONTROLS

4-Port AutoCal SetupFigure 9-3 shows the connections necessary for the operation of the4-Port AutoCal module. These include:qPower supply (connect the

Strona 264 - 6-6 DATA DISPLAY

The RF cable connections will be dependent on the type of calibrationbeing done and on the port assignments. By default, the connectionsare done as fo

Strona 265 - DISPLAYS AND DISK OUTPUT

Manual ControlWhile rarely needed, the standards within the AutoCal module can bemanually selected under:Utility/Diagnostics/AutoCal AssuranceThe indi

Strona 266 - 6-7 HARD COPY AND DISK

9-9 PIN DEPTHSPECIFICATIONSThe depth of the center pin on connectors is a critical specification,which if not met, can cause damage to mating connecto

Strona 267

AUTOCAL PIN DEPTH SPECIFICATIONSMS462XX OM 9-17/9-18Example 1:FEMALE MASTER GAUGE BLOCK (protrusion)Desired nominal value: 0.2070Case1 Actual value of

Strona 268 - AND DISK OUTPUT DISPLAYS

Inappropriate setting of the Default Gateway IP Address will causethe Scorpion system to appear to be locked up at start up. Theinstrument will appear

Strona 270

Table of Contents10-1 INTRODUCTION...10-310-2 CONVENTIONS ...10-310-3 REQUIRED EQUIPME

Strona 271 - Measurement Calibration

This equipment can not be repaired by the operator. DO NOT at-tempt to remove the equipment covers or to disassemble internalcomponents. Only qualifie

Strona 272

Chapter 10Operational CheckoutProcedures10-1 INTRODUCTION This chapter provides quick operational checkout procedures that maybe used to ensure that t

Strona 273

10-4 SELF TEST Perform an instrument self test to ensure that the MS462XX isoperating properly.Step 1. Press Power (top left) to On.Step 2. Press Util

Strona 274 - DISCUSSION CALIBRATION

10-5 MS462XA/BNON-RATIOPARAMETERS TESTThis test verifies that each individual receiver channel operatesproperly. Measurement calibration of the system

Strona 275 - CALIBRATION DISCUSSION

MS462XA/B NON-RATIO OPERATIONAL CHECKOUTPARAMETERS TEST PROCEDURES10-6 MS462XX OMFigure 10-1. Non-Ratioed Parameter Waveform

Strona 276

Step 11. For MS462XA: Press Ch1 (top left) and set resolutionfor 5 dB (use “5” and “X1” Data Entry keys); pressCh3 and set resolution for 5 dB.For MS4

Strona 277

Step 18. After at least one sweep has occurred, verify thatthe maximum-value to minimum-value amplitudeslope (Figure 10-1, page 10-6) meets the specif

Strona 278

3-Port VNMSsStep 20. Connect a Short to Port 3.Step 21. Press Display (top left).Step 22.Press DISPLAY MODE/DUAL CHANNELS 1&3.Step 23. Press Meas

Strona 279

the kit). Orient the handles such that the slotted opening in thehandle mounting plate (Figure 2-1) is toward the bottom of the unit.Step 6. For Optio

Strona 280

10-6 MS462XA/B HIGHLEVEL NOISE TESTThe following test verifies that the high-level signal noise in theMS462XX will not significantly affect the accura

Strona 281

Step 10. For MS462XB, press Meas (top left) then Ch 1(middle left). Select S12.Step 11.Press Display (middle left), TRACE MEMORY/VIEW DATA.Step 12.Aft

Strona 282

Step 16. Press Hard Copy key (top left).Step 17.Select OUTPUT DEVICE then DISK FILE.Step 18.Press RETURN.Step 19.Select OUTPUT SETUP then FLOPPY DISK.

Strona 283 - 7-4 MEASUREMENT

Step 27. Enter the following formula in the cell that followsthe last data point (top left).SORT(SUMSQ(X1:X401)/401)Where X is the column letter in th

Strona 284 - 7-14 MS462XX OM

MS462XA/B HIGH OPERATIONAL CHECKOUTLEVEL NOISE TEST PROCEDURES10-14 MS462XX OMFigure 10-2. Notepad File ExamplesHeader text is shown highlighted.Heade

Strona 285 - MS462XX OM 7-15

OPERATIONAL CHECKOUT MS462XA/B HIGHPROCEDURES LEVEL NOISE TESTMS462XX OM 10-15Figure 10-3. Excel Dialog Boxes.018Select “Delimited” in the first Excel

Strona 286 - 7-16 MS462XX OM

Step 33. Press Meas (top left).Step 34.For MS462XA, press S11.For MS462XB, press Ch 1 then S11; press Ch 3 thenS22 (middle left).Step 35. Connect a Sh

Strona 287 - MS462XX OM 7-17

10-7 MS462XC SYSTEMDYNAMIC RANGETESTThis test verifies the System Dynamic Range of the test channels ofthe MS462XC receiver.Test Setup Set up the equi

Strona 288 - 7-18 MS462XX OM

Step 7. Select:SOURCE 1 SETUPFLAT TEST PORT POWER CALStep 8. Change POWER TARGET to -10 dBm.Step 9. Zero and calibrate the power meter. Then connectth

Strona 289 - CALIBRATION (OSL) CALIBRATION

Step 22. Allow one full sweep to occur, then press Hold key.Step 23. Verify that the trace is below the REFERENCEVALUE.Step 24. Press Meas key and sel

Strona 290 - (OSL) CALIBRATION CALIBRATION

Use a Suitable ContainerObtain a corrugated cardboard carton with a275-pound test strength. This carton should have in-side dimensions of no less than

Strona 292

Table of Contents11-1 INTRODUCTION...11-311-2 PURPOSE...11-311-3 KIT CONTENTS ...

Strona 294 - 7-24 MS462XX OM

Chapter 11Calibration Kits11-1 INTRODUCTION This chapter provides illustrations and contents for the Models 3750R,3751R, and 3753R Calibration Kits.11

Strona 295 - MS462XX OM 7-25

Model 3750R 3.5 mm Calibration Kit The 3.5 mm Connector kit (Figure 11-1) includes thefollowing items:Index 1. 23LF50 Female ShortIndex 2. 23L50 Male

Strona 296 - 7-7 FOUR-PORT

CALIBRATION KITKITS CONTENTSMS462XX OM 11-5TERMINATION28LF50LFSER. XXXXXXSHORT23LF50SER. XXXXXXTERMINATION28L50LFSER. XXXXXXOPEN24LF50SER. XXXXXXTERMI

Strona 297 - 7-8 LRL/LRM CALIBRATION

Model 3751R GPC-7 Calibration Kit The GPC-7 Connector kit (Figure 11-2) includes thefollowing items:Index 1. 23A50 Male ShortIndex 2. 24A50 Female Ope

Strona 298 - CALIBRATION CALIBRATION

CALIBRATION KITKITS CONTENTSMS462XX OM 11-750 TERMINATION28A50LFS/N. XXXXXX23A50SHORTSER. XXXXXX50 TERMINATION28A50LFS/N. XXXXXX24A50OPENSER. XXXXXX12

Strona 299 - MEASUREMENT LRL/LRM

Model 3753R Type N Calibration Kit The Type N connector kit (Figure 11-3) includes thefollowing items:Index 1. 23NF50 Female ShortIndex 2. 24NF50 Fema

Strona 300

CALIBRATION KITKITS CONTENTSMS462XX OM 11-950 TERMINATION28N50LFSER. XXXXXX23N50SHORTSER. XXXXXX50 TERMINATION28N50LFSER. XXXXXX23N50OPENSER. XXXXXX50

Strona 301

2-10 SERVICE CENTERS A list of Anritsu worldwide Service Centers is provided in Table 2-1.INSTALLATION SERVICECENTERSMS462XX OM 2-9/2-10UNITED STATESA

Strona 302

11-4 PRECAUTIONS FORUSING CONNECTORSThe following are precautionary notes related to the use of connectors.For specific information on setting pin dep

Strona 303

MechanicalShockRF components are designed to withstand years ofnormal bench handling. However, do not drop or oth-erwise treat them roughly. They are

Strona 304 - 7-9 TRM CALIBRATION

11-5 CLEANINGINSTRUCTIONSConnector interfaces—especially the outer conductors on the GPC-7and SMA connectors—should be kept clean and free of dirt and

Strona 305

CALIBRATION CLEANINGKITS INSTRUCTIONSMS462XX OM 11-13/11-14D o N O T u s e I n d u s t r i a l S o l v e n t s o r W a t e r o n c o n

Strona 307

Table of ContentsA-1 INTRODUCTION ...A-3A-2 DESCRIPTION ...A-3A-3 BASIC FUNCTIONS ...

Strona 309 - Time Domain

Appendix ASequence OperationA-1 INTRODUCTION This appendix provides description of the Seq (Sequence) key functionand procedures for its use.A-2 DESCR

Strona 310

limited to approximately 500 statements. A statement consists of akeyword (such as, SAVE AUTOCAL), or a keyword and a value (suchas, START 1.2 GHz), a

Strona 311

SEQUENCE SEQUENCEOPERATIONS CREATIONMS462XX OM A-5Function DescriptionLoad A sequence may be loaded to memory from a floppy disk or hard disk.Delete A

Strona 313 - DOMAIN MEASUREMENTS

A-7 SEQUENCE CREATION For this example, you will create a program for Sequence 1. You willbe instructed to enter the parameters to setup the measureme

Strona 314

SEQUENCE SEQUENCEOPERATIONS MODIFICATIONMS462XX OM A-7A SEQUENCE OF KEYSTROKES CAN BE STORED IN MEMORY OR DISKFOR LATER PLAYBACK, THUS SAVING TIME ON

Strona 315

A-8 SEQUENCEMODIFICATIONModifying a sequence is described below.Delete a StatementStep 1. Press Seq/EDIT SEQUENCE/SEQUENCE 1.Step 2. Using the rotary

Strona 316 - 8-3 OPERATING TIME

A-9 SEQUENCE EXECUTION There are two types of sequence execution: regular and auto. Aregular execution is initiated by pressing the Seq key and choosi

Strona 317 - DOMAIN TIME DOMAIN

A-10 DISPLAY MENUDURING A SEQUENCEEXECUTIONThis function turns on the display of menu transitions in the sequencelist, and also causes menu transition

Strona 318 - TIME DOMAIN DOMAIN

A-11 SEQUENCE REMOVAL This example illustrates a sequence being removed from memory.Step 1. Press Seq/UTILITY.Step 2. Press DELETE SEQUENCE/SEQUENCE 2

Strona 319

A-14 MULTIPLE SEQUENCECASCADINGCascading multiple sequences is done by entering an EXECUTESEQUENCE statement in a sequence. The following steps illust

Strona 320 - WINDOWING TIME

A-15 SUB-SEQUENCECALLING BY LOOPCOUNTERThe loop counter enables repetition of a sequence multiple times. Thisexample creates a sequence that sets the

Strona 321 - TIME GATING

A-16 SUB-SEQUENCECALLING BY LIMITTESTINGA sequence can use the result of a limit test to jump to anothersequence or to restart itself. A designated se

Strona 322 - GATING TIME

Step 6. Enter “1” from the keypad and press the X1 key.Step 7. Press then the following soft keys:TEST LIMITSBEEP ON FAILURE ONLIMIT TESTING ONStep 8.

Strona 323 - Chapter 9

Table of Contents3-1 INTRODUCTION ...3-33-2 GENERAL DESCRIPTION ...3-3Source Module or Modu

Strona 324 - Calibration Coefficients

A-17 LOOP COUNTEREMBEDDED IN A DATAFILE NAMEIn a sequence, the loop counter value may be embedded in a data filename. This feature allows separate dat

Strona 325

Step 7. Press Seq then the following soft keys:CONTROL STATEMENTCONDITION STATEMENTLOOP COUNTER - 1IF LOOP COUNTER<>0SELECT EXISTING SEQUENCESEQ

Strona 326 - 9-3 CALIBRATIONS

A-18 TTL OUTPUT FORCONTROLLINGPERIPHERALSOn the parallel port, there are eight TTL compatible output lines thatcan be used for controlling external pe

Strona 327 - AUTOCAL DEFINITIONS

A-19 TTL INPUT CONTROLSTATEMENTOn the parallel port, there are five TTL compatible input lines thatcan be used for controlling the flow of a sequence.

Strona 328 - PHYSICAL SETUP AUTOCAL

A-20 BNC TTL OUTPUTCONTROLSTATEMENTA BNC TTL output signal can be used to control the test environment.On a two-port instrument, the driven port has t

Strona 329 - 9-7 USING AUTOCAL

A-21 GOSUB SEQUENCE The GOSUB sequence capability allows treatment of a sequence as asubroutine. After the subroutine sequence has completed, executio

Strona 330 - USING AUTOCAL AUTOCAL

A-22 NESTING GOSUBSEQUENCESNesting sub-sequences is another method of cascading sequences. Thefollowing example illustrates this.Create Sequence 1 to

Strona 331 - AUTOCAL USING AUTOCAL

Create Sequence 3 (listing appears at left)Step 15.Press Seq/EDIT SEQUENCE/SEQUENCE 3.Step 16.Press Display/AUTOSCALE.Step 17.Press Seq/CONTROL STATEM

Strona 332 - 9-8 4-PORT AUTOCAL

A-23 OPERATORMESSAGESAn operator message can be embedded in a sequence to inform theinstrument’s operator to take certain actions. An operator message

Strona 333 - AUTOCAL 4-PORT AUTOCAL

Step 2. Press Seq then the following soft keys:CONTROL STATEMENTMESSAGESEQUENCE COMMENTStep 3. Using an external keyboard or the message editingmenu,

Strona 334 - 4-PORT AUTOCAL AUTOCAL

WARRANTYThe Anritsu product(s) listed on the title page is (are) warranted against defects in materials andworkmanship for three years from the date o

Strona 336

A-24 DUMP A SEQUENCEIN ASCII TO DISKTo dump a sequence in a readable ASCII code to disk, proceed asfollows:Step 1. Press Seq then the following soft k

Strona 337

Table of ContentsB-1 INTRODUCTION ...B-3B-2 MS462XX REAR PANEL ...B-3B-3 CONNECTOR PINOUT

Strona 339 - MS462XX OM 9-17/9-18

Appendix BRear Panel ConnectorsB-1 INTRODUCTION This appendix provides descriptions and pinout diagrams for theMS462XX rear panel connectors .B-2 MS46

Strona 340

MS462XX REAR REAR PANELPANEL CONNECTORSB-4 MS462XX OMNoiseInNoise+28VNO OPERATOR SERVICE-ABLE PARTS INSIDE.REFER SERVICING TOQUALIFIED PERSONNEL.WARNI

Strona 341 - Procedures

1.Noise +28: Outputs pulsed 28V supply to drive the noise source.2.Noise In: Routes Noise Source energy to the test set.3.Line Voltage Input: Three-pr

Strona 342

13.SCSI-2: Provides for connecting an external SCSI-2 hard disk drive (Option 4). Figure B-7 provides apinout diagram.14.LPT Line Printer: 25-pin conn

Strona 343

REAR PANEL MS462XX REARCONNECTORS PANELMS462XX OM B-7PIN NAME DESCRIPTION1-4 DIO 1 thru DIO 4 Data Input/Output. Bits are HIGH with the data is logica

Strona 344 - TEST PROCEDURES

MS462XX REAR REAR PANELPANEL CONNECTORSB-8 MS462XX OMPIN NAME DESCRIPTION1 STROBEPrinter Strobe. A low-true pulse that tells the printer valid data ha

Strona 345 - 10-5 MS462XA/B

REAR PANEL MS462XX REARCONNECTORS PANELMS462XX OM B-9PIN NAME DESCRIPTION1 Limit 1-4 Rtn Return for the Channel 1-4 limit signal2 (TTL Out 0)* Channel

Strona 346 - PARAMETERS TEST PROCEDURES

Chapter 3Network Analyzers,A Primer3-1 INTRODUCTION This section provides front panel operating and measurementapplication information and data. It in

Strona 347 - PROCEDURES PARAMETERS TEST

MS462XX REAR REAR PANELPANEL CONNECTORSB-10 MS462XX OMPIN NAME DESCRIPTION1 Red Red signal2 Green Green signal3 Blue Blue signal4 Not Used5 Not Used6

Strona 348

REAR PANEL MS462XX REARCONNECTORS PANELMS462XX OM B-11PIN DESCRIPTION1 Gnd2 Collision +3 Transmit +4 Ground5 Receive +6 Gnd7 N.C.8 Gnd9 Collision -10

Strona 349

MS462XX REAR REAR PANELPANEL CONNECTORSB-12 MS462XX OM65PIN DESCRIPTION1CD2 RXD3 TXD4 DTR5 N.C.6 N.C.7RTS8 CTS9 N.C.Figure B-8. Pinout Diagram, Serial

Strona 350 - 10-6 MS462XA/B HIGH

REAR PANEL MS462XX REARCONNECTORS PANELMS462XX OM B-13/B-14PIN DESCRIPTION1 thru 25 GND26 DATA 127 DATA 228 DATA 329 DATA 430 DATA 531 DATA 632 DATA 7

Strona 352 - 10-12 MS462XX OM

The latest copy of the Vector Network Measurement System TechnicalData Sheet, Anritsu Part Number: 11410-00288, can be downloaded fromthe documents ar

Strona 354 - LEVEL NOISE TEST PROCEDURES

Table of ContentsD-1 INTRODUCTION ...D-3D-2 ORGANIZATION ...D-3D-3 CONVENTIONS...

Strona 355 - PROCEDURES LEVEL NOISE TEST

Data EntryPowerKeyboardPort 3ChannelsEnhancementStimulusSystemPort 1Port 2Probe PowerMS4623BVector Network Measurement System10 MHz-6 GHz1852963-.G/ns

Strona 356 - 10-16 MS462XX OM

Appendix DSoft-Key Menu MapsD-1 INTRODUCTION This appendix provides menu maps that document the flow of thefront panel soft-key menus.D-2 ORGANIZATION

Strona 357 - 10-7 MS462XC SYSTEM

The MS462XX is a self-contained, fully integrated measurementsystem that includes an optional time domain capability. The systemhardware consists of t

Strona 358 - DYNAMIC RANGE TEST PROCEDURES

D-3 CONVENTIONS The menu maps in this chapter use the following conventions:D-4 MENU MAPS The menu maps begin with Figure D-2, on the next page.Menu M

Strona 359 - PROCEDURES DYNAMIC RANGE TEST

MS462XX OM D-5APPENDIX D Appl Key Menu Map (1 of 3)HELP ON / OFFSOURCE SELECTION+ / - xxx.xxxxxxxxx GHzTONE 2 OFFSETxxxxDISPLAY SELECTIONSOURCE / CW R

Strona 360

D-6 MS462XX OMAppl Key Menu Map (2 of 3) APPENDIX DRETURNFREQUENCIESCLEAR ALLFREQUENCYCLEAR SELECTEDxxTO BE CLEAREDFREQUENCY NUMBERFREQUENCYINSERT NEX

Strona 361 - Calibration Kits

MS462XX OM D-7APPENDIX D Appl Key Menu Map (3 of 3)HELP ON/OFFTIME / DISTANCEDISPLAY,TIME,BANDPASSTIME,LOWPASSDOMAINX.XRELATIVE VELOCITYRETURNMOREIMPU

Strona 362

D-8 MS462XX OMMeas Key Menu Map, 2- and 3-port Instruments (1 of 2) APPENDIX DMIXED MODES-PARAMETERMOREEXT ANALOG INb2/a2 / labelS22, REFL/USER 4b1/a2

Strona 363

MS462XX OM D-9APPENDIX D Meas Key Menu Map, 4-port Instruments (2 of 2)1 ( UNITY )MOREa3 (Rc)a4 (Rd)a2 (Rb)a1 (Ra)x/xSxx / USER yNUMERATORS-PARAMETERU

Strona 364 - CONTENTS KITS

D-10 MS462XX OMDisplay Key Menu Map (1 of 3) APPENDIX DAAACSheet 2BSheet 2DSheet 2CSheet 2BSheet 2CSheet 2CSheet 2BSheet 2CSheet 2CSheet 2BSheet 2CShe

Strona 365 - MS462XX OM 11-5

MS462XX OM D-11APPENDIX D Display Key Menu Map (2 of 3)MORESEGMENTED LIMITSTEST LIMITSREALMORELIMITSDISPLAY ON / OFFREADOUT LIMITxx.xxx pULOWER LIMIT

Strona 366

D-12 MS462XX OMDisplay Key Menu Map (3 of 3) APPENDIX DARETURNxREFERENCE LINExx.xx dBREFERENCE VALUExx.xx dB / DIVRESOLUTION- LOG MAGNITUDE -SCALERETU

Strona 367 - MS462XX OM 11-7

MS462XX OM D-13APPENDIX D Marker Key Menu Map, Frequency Sweep (1 of 2)ToMarkersFUNCTIONSMARKER READOUTREADOUTMARKERSREF MODE ON / OFFDMARKERSDISPLAY

Strona 368

AnalyzerModuleThe analyzer module down-converts, receives, andinterprets the IF signal for phase and magnitudedata. It then displays the results of th

Strona 369 - MS462XX OM 11-9

D-14 MS462XX OMMarker Key Menu Map, Power Sweep (2 of 2) APPENDIX DToMarkersFUNCTIONSMARKER READOUTREADOUTMARKERSREF MODE ON / OFFDMARKERSDISPLAY ON /

Strona 370 - 11-4 PRECAUTIONS FOR

MS462XX OM D-15APPENDIX D Freq Key Menu Map (1 of 1)FreqDISCRETE FILLMARKER SWEEPxxx.xxxxxxxxx GHzC.W. MODE ON / OFFSET CENTER/SPANxxx.xxxxxxxxx GHzST

Strona 371 - KITS USING CONNECTORS

D-16 MS462XX OMPower Key Menu Map (1 of 2) APPENDIX DPowerRECEIVER CALSOURCE 4 SETUPSOURCE 3 SETUPSOURCE 2 SETUPSOURCE 1 SETUPPOWERRETURNCORRECTIONAT

Strona 372 - 11-5 CLEANING

MS462XX OM D-17APPENDIX D Power Key (Power Sweep) Menu Map (2 of 2)PowerFreqxxx.xxxxxxxxx GHzC.W. MODE ON / OFFxxx.xxxxxxxxx GHzSTOPxxx.xxxxxxxxx GHzS

Strona 373 - W A T E R

D-18 MS462XX OMSweep Key Menu Map (1 of 1) APPENDIX DSweepAUTO / MANUALSWEEP TIME MODEx.xx SSWEEP TIMESWEEP TIME ON/OFF(xxxxx REMAINING )xxxxx SWEEPST

Strona 374

MS462XX OM D-19APPENDIX D Config Key Menu Map (1 of 1)ConfigHOLDMOREDATA POINTSMULTIPLE SOURCESOURCECONFIGURATIONRETURNSOURCE 4SOURCE 3SOURCE 2SOURCE

Strona 375 - Sequence Operation

D-20 MS462XX OMSequence Key Menu Map (1 of 1) APPENDIX DSeqEDITING FUNCTIONSUTILITYTTL I/OEDITDONE SEQUENCECONTROL STATEMENTGOSUB SEQUENCEEDIT SEQUENC

Strona 376

MS462XX OM D-21APPENDIX D Avg Key Menu Map (1 of 1)AvgBANDWIDTHSELECT I.F.xxFIGURE AVERAGENOISEFIGURE AVERAGENOISE ON / OFFxx.xx % OF SWEEPSMOOTHINGSM

Strona 377

D-22 MS462XX OMUtility Key Menu Map (1 of 3) APPENDIX DUtilityRETURNNEXT PARAM PAGECHANNEL 3 & 4CHANNEL 1 & 2OPERATIONCALIBRATIONSYSTEMPARAMET

Strona 378 - A-5 KEYS OPERATE

MS462XX OM D-23APPENDIX D Utility Key Menu Map (2 of 3)ASheet 3BSheet 1CSheet 1AARETURNASSURANCEAUTO CAL( SERVICE USE ONLY )TROUBLE SHOOTING( SERVICE

Strona 379 - OPERATIONS CREATION

frequency (IF). This signal can then be measureddirectly by a tuned receiver. The tuned receiverapproach gives the system greater dynamic range.The sy

Strona 380 - CREATION OPERATIONS

D-24 MS462XX OMUtility Key Menu Map (3 of 3) APPENDIX DASheet 2ABBMOREWRITE DATA= XXXREAD DISK = XXXTRIGGER READXXX(CHANNEL NAME)DVM CHANNELA/D INPUTE

Strona 381 - OPERATIONS MODIFICATION

MS462XX OM D-25APPENDIX D Hard Copy Key Menu Map (1 of 1)CopyHardSTOP PRINTPORTRAIT/ LANDSCAPEPAGE ORIENTATIONOUTPUT OPTIONSOUTPUT HEADERSSETUPOUTPUT

Strona 382 - A-8 SEQUENCE

D-26 MS462XX OMSave-Recall Key Menu Map (1 of 1) APPENDIX DRecallSave/RECALLSAVEAND CAL DATAFRONT PANELSAVE / RECALLRETURNSAVE CALINTERNAL MEMORYCAL 1

Strona 383 - OPERATIONS EXECUTION

MS462XX OM D-27APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(1 of 17)RETURNRETURNRETURNRETURNRETURNRETURNINFORMATIONDISPLAY WAVE

Strona 384 - A-10 DISPLAY MENU

D-28 MS462XX OMCalibration Key Menu Map (Transmission & Reflection)(2 of 17) APPENDIX DAEXITSAVEAUTO CALDPage D-25APage D-15ASheet 3CalDoneTHRU UP

Strona 385 - OPERATIONS REMOVAL

MS462XX OM D-29APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(3 of 17)AUTOCALMEASURING...ABORT CALAUTOCALSAVEEXITDPage D-26APage

Strona 386 - A-14 MULTIPLE SEQUENCE

D-30 MS462XX OMCalibration Key Menu Map (Transmission & Reflection)(4 of 17) APPENDIX D*AxANEXT CAL STEPLINE TYPEMETHOD ANDCHANGE CALXXXLINE TYPE:

Strona 387 - A-15 SUB-SEQUENCE

MS462XX OM D-31APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(5 of 17)A1StandardCalibrationCoaxSTART CALTEST SIGNALSIMPEDANCEREFE

Strona 388 - A-16 SUB-SEQUENCE

D-32 MS462XX OMCalibration Key Menu Map (Transmission & Reflection)(6 of 17) APPENDIX DA2StandardCalibrationMicrostripNEXT CAL STEP+XX.XXXX mmOFFS

Strona 389 - OPERATIONS BY LIMIT TESTING

MS462XX OM D-33APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(7 of 17)BSheet 10ASheet 6NEXT CAL STEPNEXT CAL STEPXX.XXX e -12INDU

Strona 390 - A-17 LOOP COUNTER

analyzer would measure a phase difference of 0 de-grees.Now consider that we make this same measurementat 1.1 GHz. The frequency is higher by 10 perce

Strona 391 - MS462XX OM A-17

D-34 MS462XX OMCalibration Key Menu Map (Transmission & Reflection)(8 of 17) APPENDIX DNEXT CAL STEPNEXT CAL STEPXX.XXX e -12INDUCTANCEXX.XXX WIMP

Strona 392 - A-18 TTL OUTPUT FOR

MS462XX OM D-35APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(9 of 17)NEXT CAL STEPXX.XXX e -12INDUCTANCEXX.XXX WIMPEDANCEPARAMET

Strona 393 - A-19 TTL INPUT CONTROL

D-36 MS462XX OMCalibration Key Menu Map (Transmission & Reflection)(10 of 17) APPENDIX DA6LRL/LRMCalibrationCoaxA7LRL/LRMCalibrationMicrostripA8LR

Strona 394 - A-20 BNC TTL OUTPUT

MS462XX OM D-37APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(11 of 17)MEASURE PORT 1((NO) CAL EXISTS)xxxxxxxxxxxxxxPORT 2:xxxxxx

Strona 395 - OPERATIONS CONTROL STATEMENT

D-38 MS462XX OMCalibration Key Menu Map (Transmission & Reflection)(12 of 17) APPENDIX DABNEXT CAL STEPMETHODCHANGE CALXXXLINE TYPE:XXXCAL METHODP

Strona 396 - A-22 NESTING GOSUB

MS462XX OM D-39APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(13 of 17)AMEASURE DEVICES3 PORTCALIBRATIONCONNECT ATHROUGHLINEBETWE

Strona 397 - OPERATIONS SEQUENCES

D-40 MS462XX OMCalibration Key Menu Map (Transmission & Reflection)(14 of 17) APPENDIX DANEXT CAL STEPMETHODCHANGE CALUSE EXISTING 3-PORTCAL YES/N

Strona 398 - A-23 OPERATOR

MS462XX OM D-41APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(15 of 17)ATO PROCEEDPRESS <ENTER>ORDATA ON DISKTO STORE CAL&l

Strona 399 - OPERATIONS MESSAGES

D-42 MS462XX OMCalibration Key Menu Map (Transmission & Reflection)(16 of 17) APPENDIX DNEXT CAL STEPREMEASUREMEASUREMENTRESTARTMEASURING...MEASUR

Strona 400 - A-25 SAVING A SEQUENCE

MS462XX OM D-43APPENDIX D Calibration Key Menu Map (Transmission & Reflection)(17 of 17)TO PROCEEDPRESS <ENTER>ORDATA ON DISKTO STORE CAL<

Strona 401 - Rear Panel Connectors

There are two ways of accomplishing this. The mostobvious way is to insert a length of line into the ref-erence signal path to make both paths of equa

Strona 402

D-44 MS462XX OMCalibration Key Menu Map (Harmonic)(1 of 1) APPENDIX DAPage D-16BPage D-16HELP ON / OFFNONE( (NO) CAL EXISTS )AND PHASEENHANCEMENT( (NO

Strona 403

MS462XX OM D-45APPENDIX D Calibration Key Menu Map (IMD)(1 of 1)AorBPage D-16CalMeasurement TypeIMDApplPressPressSelectHELP ON / OFFPERFORM CAL(( NO)

Strona 404 - B-4 MS462XX OM

Calibration Key Menu Map (Noise Figure)(1 of 1) APPENDIX DD-46 MS462XX OMAPage D-15NOISE FIGUREHELP ON / OFFWITH 12-TERM*NOISE FIGURE*Not allowed with

Strona 405 - CONNECTORS PANEL

APPENDIX D Calibration Key Menu Map (Gain Compression)(1 of 1)MS462XX OM D-47CalMeasurement TypeGAIN COMPRESSIONSWEPT FREQUENCYthenREFLECTIONTRANSMISS

Strona 406 - PANEL CONNECTORS

Calibration Key Menu Map (Frequency Group Delay)(1 of 1) APPENDIX DD-48 MS462XX OMFREQUENCYTRANSLATIONGROUP DELAYHELP ON / OFFPERFORM CALAPPLY CAL ON/

Strona 407

Subject Index#10MHzRefIN...B-520 dB Expanded Smith Chart ...6-112-Port Calibration...7-193 dB Compressed Smith Chart

Strona 408

DData DisplayActive Channel Selection...6-16Analog Instrument Status ...6-15Disk Output...6-19Frequency...6-12Freque

Strona 409

Fourier Transform ...8-3Four-Port Calibration...7-26FREQ Key Menus ...4-62frequency domain...8-8Fre

Strona 410

MarkersDisplay of ...6-18Reference Position ...6-14Markers/Limits Keys ...4-3MEAS Key Menus...4-20Measur

Strona 411

Purpose, of Calibration Kits ...11-3RRack Mount (Option 1) Installation ...2-6Rear PanelConnector Key ...B-4TOB-13Diagrams...

Strona 412

NetworkAnalyzerMeasurementsNow let us consider measuring the DUT. Consider atwo port device; that is, a device with a connector oneach end. What measu

Strona 413

TTabular Printout ...6-18Teflon Tuning Washers ...11-10Test Ports, Discussion ...7-3Test Set Module...

Strona 414

There are several ways in which all the informationcan be displayed on one trace. One method is a polardisplay (Figure 3-17). The radial parameter (di

Strona 415 - Specifications

Another important parameter we can measure whenphase information is available is group delay. Inlinear devices, the phase change through the DUT islin

Strona 417 - Soft-Key Menu Maps

Summary A network analyzer is similar to a scalar networkanalyzer. The major difference it that it adds the ca-pability for measuring phase as well as

Strona 418 - 10 MHz-6 GHz

Table of Contents4-1 INTRODUCTION ...4-34-2 FRONT PANEL KEY DESCRIPTIONS ...4-34-3 SOFT-KEY MENUS,

Strona 419

Data EntryPowerKeyboardPort 3ChannelsEnhancementStimulusSystemPort 1Port 2Probe PowerMS4623BVector Network Measurement System10 MHz-6 GHz1852963-.G/ns

Strona 420 - D-4 MS462XX OM

Chapter 4Front Panel Operation4-1 INTRODUCTION This chapter describes the front panel keys, controls, menus, andrelated GPIB mnemonics. The chapter is

Strona 421 - MS462XX OM D-5

Index 12 Stimulus KeysFreq: Displays the first in a series of menus that provide fre-quency-sweep control and let you trigger I.F. calibration measure

Strona 422 - D-6 MS462XX OM

Index 16 System KeysPrint: Tells the printers, plotters or disk files to start output basedon the current selections.Hard Copy: Displays the first in

Strona 423 - MS462XX OM D-7

b. GPIB Mode: The key returns the instrument to local (front panel)control, unless the controller has sent a local lockout message (LLO)over the bus.I

Strona 424 - D-8 MS462XX OM

4-4 APPL KEY MENUS This key provides entry to measurement application functions.Menu Name Function Description GPIB MnemonicAPPLICATIONSMEASUREMENT TY

Strona 425 - MS462XX OM D-9

Menu Name Function Description GPIB MnemonicD.C. TERMSETUP FORLOWPASSPROCESSINGAUTO EXTRAPOLATE Sets the D.C. term to a value determined byextrapolati

Strona 426 - D-10 MS462XX OM

Menu Name Function Description GPIB MnemonicDOMAIN2RANGE SETUP Changes the soft-key display to a DISPLAYSETUP menu.NoneGATE SETUP Changes the soft-key

Strona 428 - D-12 MS462XX OM

Menu Name Function Description GPIB MnemonicHARMONICSWEEP MODESOURCE/CW RCVRSweeps the source frequency or receiverfrequency.APPSWPC; APPSWPS;APPSWPX?

Strona 429 - MS462XX OM D-13

Menu Name Function Description GPIB MnemonicIMD SOURCESELECTIONSELECTOR MODEDEFINE/APPLYSwitches between DEFINE and APPLY setupmodes. The parameter ch

Strona 430 - D-14 MS462XX OM

Menu Name Function Description GPIB MnemonicLOWPASSDISTANCEDISPLAY SETUPSTARTXXX.XXX mmSets the start distance of the display. GST; GST?STOPXXX.XXX mm

Strona 431 - MS462XX OM D-15

Menu Name Function Description GPIB MnemonicMEASUREMENTTYPETRANSMISSION ANDREFLECTIONChanges the soft-key display to theTRANSMISSION AND REFLECTION me

Strona 432 - D-16 MS462XX OM

Menu Name Function Description GPIB MnemonicMIXERLO/RECEIVERSETUPLO IS SOURCE(2-4)XSelects the LO source. APPLORS2-APPLORS4;APPLORSX?LO OFFSET FROM RF

Strona 433 - MS462XX OM D-17

Menu Name Function Description GPIB MnemonicMULTIPLEFREQUENCYGAINCOMPRESSIONTEST AUT Changes the soft-key display to the MULTIPLEFREQUENCY GAIN COMPRE

Strona 434 - D-18 MS462XX OM

Menu Name Function Description GPIB MnemonicNOISE FIGURESETUPNOISE SOURCEINTERNAL/EXTERNALToggles noise source between Internal andExternal.NFSRCI; SF

Strona 435 - MS462XX OM D-19

Menu Name Function Description GPIB MnemonicSWEPTFREQUENCYGAINCOMPRESSIONNOMINAL OFFSET XXX dB Sets nominal offset. NOFST; NOFST?GAIN COMPRESSIONPOINT

Strona 436 - D-20 MS462XX OM

Menu Name Function Description GPIB MnemonicSWEPT POWERSETUPSWEPT POWERFREQUENCYXXX.XXXXXX GHzSets the power sweep frequency. PSF; PSF?P START XX.XX d

Strona 437 - MS462XX OM D-21

Menu Name Function Description GPIB MnemonicWINDOW/GATESHAPE SETUPRECTANGULAR Selects a Rectangular (one-term) shape. WRT; GRTNOMINAL Selects a two-te

Strona 438 - D-22 MS462XX OM

Safety SymbolsTo prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Company uses thefollowing symbols to indicate s

Strona 439 - MS462XX OM D-23

4-5 MEAS KEY MENUS This key provides entry to measurement functions.Menu Name FunctionDescriptionGPIB MnemonicSELECTBALANCEDPORT PAIR(3 PORT MODELSWIT

Strona 440 - D-24 MS462XX OM

Menu Name FunctionDescriptionGPIB MnemonicSELECTBALANCEDPORT PAIR(2ND PAIR)(4 PORT MODELS4 PORT DEVICES)(1:2)/(2:1) Selects Ports 1:2 or 2:1 as the se

Strona 441 - MS462XX OM D-25

Menu Name FunctionDescriptionGPIB MnemonicSELECTMIXED MODES-PARAMETERSDCMODE CONVERSIONSelects SDC to be the active S-parameter. MSDC; MMSDC*SCDMODE C

Strona 442 - D-26 MS462XX OM

Menu Name FunctionDescriptionGPIB MnemonicSELECTMIXED MODES-PARAMETER5(4 PORTMODELS)SD2C1MODE CONVERSIONSelects SD2C1 to be the active S-parameter. MS

Strona 443 - MS462XX OM D-27

Menu Name FunctionDescriptionGPIB MnemonicSELECTS-PARAMETERS13, TRANS/USER 5b1/a3Selects S13 to be the active S-parameter. S13S23, TRANS/USER 6b2/a3Se

Strona 444 - D-28 MS462XX OM

Menu Name FunctionDescriptionGPIB MnemonicUSER DEFINEDS-PARAMETERDENOMINATORSxx/USERxx/yDisplays the active S-parameter and its ratio. USR1 to USR9; U

Strona 445 - MS462XX OM D-29

Menu Name FunctionDescriptionGPIB MnemonicUSER DEFINEDS-PARAMETERNUMERATORSxx/USERxx/yDisplays the active S-parameter and its ratio. USR1 to USR9; UDP

Strona 446 - D-30 MS462XX OM

Menu Name FunctionDescriptionGPIB MnemonicUSER DEFINEDS-PARAMETERNUMERATOR(4 PORTMODELS)Sxx/USERxx/yDisplays the active S-parameter and its ratio. USR

Strona 447 - MS462XX OM D-31

4-6 DISPLAY KEY MENUS This key provides entry to display functions.Menu Name Function Description GPIB MnemonicDISPLAYDISPLAY MODE Changes the soft-ke

Strona 448 - D-32 MS462XX OM

Menu Name Function Description GPIB MnemonicDISPLAY MODESINGLE CHANNEL Selects a single channel for display. You select thetype of display in menu GRA

Strona 449 - MS462XX OM D-33

Safety 2 MS462XX OMAlways refer to the operation manual when working near locationswhere the alert mark, shown on the left, is attached. If the opera-

Strona 450 - D-34 MS462XX OM

Menu Name Function Description GPIB MnemonicEDIT UPPERLIMIT SEGMENTSGRAPH TYPESEGMENT X ON/OFF Enter the segment number that you want to define,and tu

Strona 451 - MS462XX OM D-35

Menu Name Function Description GPIB MnemonicGRAPH TYPES 2POWER OUT Provides for measuring output power. Themeasurement of output power is accomplished

Strona 452 - D-36 MS462XX OM

Menu Name Function Description GPIB MnemonicREADOUT LIMIT-GRAPH TYPE- Displays the currently selected graph type. NoneUPPER LIMIT (REF)XX.XX dBSets th

Strona 453 - MS462XX OM D-37

Menu Name Function Description GPIB MnemonicREFERENCEPLANE PERPORTPORT X Select the port number. RPPORTNUM;RPPORTNUM?AUTO Changes the S-parameter bein

Strona 454 - D-38 MS462XX OM

Menu Name Function Description GPIB MnemonicSCALEIMAGINARYRESOLUTION XX.XXXU/DIVSets the resolution for the vertical axis of the activechannel display

Strona 455 - MS462XX OM D-39

Menu Name Function Description GPIB MnemonicSCALE LINEARMAG & PHASE(CAL)-LINEAR MAG-NoneRESOLUTIONXX.XXX U/DIVSets the resolution for the vertical

Strona 456 - D-40 MS462XX OM

Menu Name Function Description GPIB MnemonicSCALE LOG MAG& PHASE-LOG MAG-NoneRESOLUTIONXX.XX dB/DIVSets the resolution for the vertical axis of th

Strona 457 - MS462XX OM D-41

Menu Name Function Description GPIB MnemonicSCALE LOGMAGNITUDERESOLUTIONXX.XX dB/DIVSets the resolution for the vertical axis of the activechannel dis

Strona 458 - D-42 MS462XX OM

Menu Name Function Description GPIB MnemonicSCALE POWEROUT (CAL)RESOLUTIONXX.XX dB/DIVSets the resolution for the vertical axis of the activechannel d

Strona 459 - MS462XX OM D-43

Menu Name Function Description GPIB MnemonicSCALE REAL &IMAGINARY CAL)-REAL-NoneRESOLUTIONXX.XXX U/DIVSets the resolution for the vertical axis of

Strona 460 - D-44 MS462XX OM

Anritsu products comply with the immunity requirements defined by the European Immunity StandardEN 61000-4-2:1995/EN50082-1:1997-4kV CD, 8kV AD for st

Strona 461 - MS462XX OM D-45

Menu Name Function Description GPIB MnemonicSCALE SWRRESOLUTIONXX.XXX U/DIVSets the resolution for the vertical axis of the activechannel displayed gr

Strona 462 - D-46 MS462XX OM

Menu Name Function Description GPIB MnemonicSELECT FILE TOREADFILE 1 Recalls the data stored in file number 1 NoneFILE 2 Recalls the data stored in fi

Strona 463 - MS462XX OM D-47

Menu Name Function Description GPIB MnemonicSET DIELECTRICAIR(1.000649)Calculates reference delay based on the dielectricconstant of air (1.000649).DI

Strona 464 - D-48 MS462XX OM

Menu Name Function Description GPIB MnemonicSINGLE LIMITSIMAGINARY-IMAGINARY- NoneUPPER LIMITXX.XXXXXX U ON/OFFTurns the Upper Limit line on or off fo

Strona 465 - Subject Index

Menu Name Function Description GPIB MnemonicSINGLE LIMITSLINEAR MAG &PHASE-LINEAR MAG- NoneUPPER LIMITXX.XXXXXX U ON/OFFTurns the Upper Limit line

Strona 466 - D SUBJECT INDEX

Menu Name Function Description GPIB MnemonicSINGLE LIMITSLOG MAGNITUDEUPPER LIMITXX.XXXXXX dBON/OFFTurns the Upper Limit line on or off for the active

Strona 467 - SUBJECT INDEX G

Menu Name Function Description GPIB MnemonicSINGLE LIMITSLOG POLAR-LOG POLAR- NoneUPPER LIMITXX.XXXXXX dBON/OFFTurns the Upper Limit line on or off fo

Strona 468 - N SUBJECT INDEX

Menu Name Function Description GPIB MnemonicSINGLE LIMITSREAL-REAL- NoneUPPER LIMITXX.XXXXXX U ON/OFFTurns the Upper Limit line on or off for the acti

Strona 469 - SUBJECT INDEX R

Menu Name Function Description GPIB MnemonicSINGLE LIMITSSMITH CHARTUPPER LIMITXX.XXXXXX UON/OFFTurns the upper limit on or off for the active channel

Strona 470 - T SUBJECT INDEX

Menu Name Function Description GPIB MnemonicTRACE MEMORYCHANNEL XVIEW: NoneDATA Displays measured data; that is, the data presentlybeing taken.DATMEMO